Model 400E Ozone Analyzer
Instruction Manual
Specifications, Agency Approvals, and
Warranty
P/N 04316 Rev: B
17
2.
SPECIFICATIONS, AGENCY APPROVALS,
AND WARRANTY
2.1.
Specifications
Table 2-1: Model 400E Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
Min: 0-100 PPB
Max: 0-10,000 PPB
Measurement Units
ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
Zero Noise
< 0.3 ppb RMS (EPA Definition)
Span Noise
< 0.5% of reading above 100 PPB (EPA Definition)
Lower Detectable Limit
< 0.6 PPB (EPA Definition)
Zero Drift (24 hours)
< 1.0 ppb (at constant temperature and voltage)
Zero Drift (7 days)
< 1.0 ppb (at constant temperature and voltage)
Span Drift (24 hours)
< 1% of reading (at constant temperature and voltage)
Span Drift (7 days)
< 1% of reading (at constant temperature and voltage)
Linearity
< 1% of full scale
Precision
< 0.5% of reading (EPA Definition)
Lag Time
< 10 sec (EPA Definition)
Rise/Fall Time
< 20 sec to 95% (EPA Definition)
Sample Flow Rate
800
±
80 cc/min
Temperature Range
5 - 40
°
C
Humidity Range
0-90% RH, Non-Condensing
Pressure Range
25 – 31 “Hg-A
Altitude Range
0-2000m
Temp Coefficient
< 0.05% per deg C
Voltage Coefficient
< 0.05% per Volt AC (RMS) over range of nominal
±
10%
Dimensions (H x W x D)
7” x 17” x 23.5”
Weight
30.6lbs. (13.8Kg) with IZS Option
AC Power
100V 50/60Hz (3.25A), 115V 60Hz (3.0A), 220 – 240 V 50/60 Hz (2.5A)
Environmental Conditions
Installation Category (Over voltage Category) II Pollution Degree 2
Analog Outputs
Four (4) Outputs, Three (3) defined
Analog Output Ranges
All Outputs: 100 mV, 1 V, 5 V, 10 V
Two concentration outputs convertible to 4-20 mA isolated current loop
All Ranges with 5% Under/Over Range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage
Status Outputs
8 Status outputs from opto-isolators
Control Inputs
6 Control Inputs, 3 defined, 3 spare
Serial I/O
COM1: RS-232; COM2: RS-232 or RS-485
Baud Rate : 300 – 115200
Certifications
USEPA: Equivalent Method Number EQOA-0992-087
CE Mark
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...