
Model 400E Ozone Analyzer
Instruction Manual
CALIBRATION PROCEDURES
P/N 04316 Rev: B
173
STEP THREE:
Zero and Span calibrations using the Zero/Span Valve option are
similar to that described in Section 7.1, except that:
Zero Air and Span Gas is supplied to the analyzer through the Zero Gas and
Span Gas inlets rather than through the Sample inlet.
The Zero And Cal operations are initiated directly and independently with
dedicated keys (
CALZ
&
CALS
).
SAMPLE
RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 500.0 PPB O3 =XXX.X
< TST TST >
ZERO
CONC
Press
EXIT
to Return to the
Main
SAMPLE
Display
M-P CAL
RANGE = 500.0 PPB O3 =XXX.X
< TST TST >
ENTR
CONC
EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
SPAN
button
is displayed
WARNING!
Pressing
ENTR
changes the
calibration of the instrument.
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
ZERO
button
is displayed
If either the
ZERO
or
SPAN
buttons
fail to appear see
Section 9 for
troubleshooting
tips.
Analyzer enters
ZERO CAL
M ode
See Table 5-1 for Z/S
Valve States during
this operating mode
EXIT
halts the calibration
process and returns the
unit to
SAMPLE
Mode.
ZERO CAL M
RANGE = 50.0 PP
CO = 0.00
< TST TST > ZERO CONC
EXIT
SAMPLE
RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 500.0 PPB O3 =XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL
RANGE = 500.0 PPB O3 =XXX.X
< TST TST >
ENTR
CONC
EXIT
WARNING!
Pressing
ENTR
changes the
calibration of the instrument.
EXIT
halts the calibration
process and returns the
unit to
SAMPLE
Mode.
Analyzer enters
SPAN CAL
Mode
See Table 5 -1 for Z/S
Valve States during
this operating mode
ZERO CAL M
RANGE = 500.0 PPB O3 =XXX.X
< TST TST >
SPAN
CONC
EXIT
Press
ZERO
only if
performing a ZERO
CALIBRATION
If performing a ZERO
CHECK, See Section 7.5.
Press
SPAN
only if
performing a SPAN
CALIBRATION
If performing a SPAN
CHECK, See Section 7.5.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...