Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B
157
REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF
If the instrument is powered off in the middle of a
REPORT PERIOD
, the samples
accumulated so far during that period are lost. Once the instrument is turned back
on, the iDAS restarts taking samples and temporarily them in volatile memory as
part of the
REPORT PERIOD
currently active at the time of restart. At the end of
this
REPORT PERIOD
only the sample readings taken since the instrument was
turned back on will be included in any
AVG
,
MIN
or
MAX
calculation. Also, the
STORE NUM. SAMPLES
feature will report the number of sample readings taken
since the instrument was restarted.
To define the
REPORT PERIOD
, follow the instruction shown in section 6.11.2.2
then press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu.
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
Press
SET>
key until you reach
REPORT PERIOD
…
SETUP X.X
REPORT PERIODD:DAYS:0
0 0 0
ENTR EXIT
ENTR
accepts the new string and
returns to the previous menu.
EXIT
ignores the new string and
returns to the previous menu.
Press keys to set hours
between reports in the format :
HH:MM (max: 23:59). This is a
24 hour clock . PM hours are 13
thru 23, midnight is 00:00.
Example 2:15 PM = 14:15
Set the number of days
between reports (0-366).
IIf at any time an illegal entry is selected (e.g., days > 366)
the ENTR key will disappear from the display.
SETUP X.X
REPORT PERIODD:TIME:01:01
0 0 0 5
ENTR EXIT
SETUP X.X
REPORT PERIOD:000:00:05
<SET SET>
EDIT
EXIT
Use the PREV and NEXT
keys to scroll to the data
channel to be edited.
From the
DATA ACQUISITION
menu
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...