Table 6-2: Test Functions Defined (continued)
Parameter
Display Title
Units
Meaning
O3GENDRIVE
O3 DRIVE
(1)
mV
The Drive voltage used to control the intensity of
the O
3
generator UV Lamp.
(1)
SAMPPRESS
PRES
In-Hg-A
The absolute pressure of the Sample Gas as
measured by a solid state pressure sensor.
SAMPFLOW
SAMP FL
cc/min
Sample Gas mass flow rate as measured by the
Flow Sensor located between the Optical Bench
and the Sample Pump.
SAMPTEMP
SAMPLE TEMP
°
C
The Temperature of the gas inside the Sample
Chamber.
PHOTOLTEMP
PHOTO LAMP
°
C
The Temperature of the UV Lamp in the Optical
Bench.
O3SCRUBTEM
P
O3 SCRUB
(3)
°
C
The current temperature of the Metal Wool
Scrubber.
(3)
O3GENTEMP
O3 GEN TMP
(1)
°
C
The Temperature of the UV Lamp in the O
3
Generator.
(1)
BOXTEMP
BOX TEMP
°
C
The temperature inside the analyzer chassis.
SLOPE
SLOPE
- -
The Slope of the instrument as calculated during
the last calibration activity.
When the unit is set for
SINGLE
or
DUAL
Range
mode, this is the
SLOPE
of
RANGE1
.
When the unit is set for
AUTO
Range mode, this
is the
SLOPE
of the current range.
OFFSET
OFFSET
PPB
The Offset of the instrument as calculated during
the last calibration activity.
When the unit is set for
SINGLE
or
DUAL
Range
mode, this is the
OFFSET
of
RANGE1
.
TESTCHAN
TEST
(4)
mV
Displays the signal level of whatever Test
function is currently being output by the Analog
Output Channel
A4
.
(4)
CLOCKTIME
TIME
HH:MM:SS
The current time. This is used to create a time
stamp on iDAS readings, and by the AutoCal
feature to trigger calibration events.
(1)
Only appears if IZS option is installed.
(2)
Only appears if IZS Reference Sensor option is installed.
(3)
Only appears if Metal Wool Scrubber option is installed.
(4)
Only appears if Analog Output A4 is activ ely reporting a Test Function.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...