Model 400E Ozone Analyzer
Instruction Manual
Getting Started
P/N 04316 Rev: B
37
To view the current values of these parameters press the following key sequence on
the analyzer’s front panel:
RANGE
STABIL
O3 MEAS
O3 REF
O3 GEN
2
O3 DRIVE
1
PRES
SAMP FL
SAMPLE TEMP
PHOTO LAMP
O3 SCRUB
3
O3 GEN TMP
1
BOX TEMP
SLOPE
TEST
4
OFFSET
TIME
SAMPLE
RANGE = 50.0 MGM CO =XX.XX
< TST
TST >
CAL
SETUP
1
Only appears if IZS option is
installed.
2
Only appears if IZS Reference
Sensor option is installed.
3
Only appears if Metal Wool
Scrubber option is installed
4
Only appears if Analog Output
A4 is actively reporting a Test
Function
Remember until the unit has completed its warm up these parameters may not
have stabilized.
3.3.
Initial Calibration Procedure
The next task is to calibrate the analyzer.
NOTE
If you are using the M400E for EPA monitoring, only the
calibration method described in Section 8 should be used.
NOTE
The detection of O
3
is subject to interference from a number of sources including,
SO
2
, NO
2
, NO, H
2
O AND aromatic hydrocarbon meta-xylene and Mercury vapor. The
Model 400E successfully rejects interference from all of these except Mercury Vapor.
Mercury Vapor absorbs radiation in the same wave-length band as O
3
but so much
more efficiently that its presence will render the analyzer useless for detecting O
3
.
If the Model 400E is installed in an environment where the presence of Mercury vapor
is suspected, specific tests should be conducted to reveal the amount of interference
and steps should be taken to remove the Mercury Vapor from the Sample Gas before
it enters the analyzer. For More information see Section 10.1.4.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...