CALIBRATION PROCEDURES
Model 400E Ozone Analyzer
Instruction Manual
180
P/N 04316 Rev: B
The following example sets Sequence #2 to do a Zero-Span Calibration every other
day starting at 1 Am on September 4, 2001, lasting 15 minutes, without calibration.
This will start ½ hour later each iteration.
Mode and Attribute
Value
Comment
Sequence
2
Define Sequence #2
Mode
ZERO-HI
Select Ze ro and Span Mode
Timer Enable
ON
Enable the timer
Starting Date
Sept. 4, 2001
Start after Sept 4, 2001
Starting Time
01:00
First Span starts at 1:00AM
Delta Days
2
Do Sequence #2 every other day
Delta Time
00:30
Do Sequence #2 ½ hr later each day
Duration
15.0
Operate Span valve for 15 min
Calibrate
NO
Do not calibrate at end of Sequence
NOTE
The programmed STARTING_TIME must be a minimum of
5 minutes later than the real time clock (See Section 6.3.5)
for setting real time clock.
NOTE
Avoid setting two or more sequences at the same time of the
day. Any new sequence which is initiated whether from a timer,
the COM ports, or the contact closure inputs will override any
sequence which is in progress.
NOTE
The CALIBRATE attribute must always be set to NO on analyzers
with IZS Options installed and functioning.
Calibrations should ONLY be performed using external sources of
Zero Air and Span Gas whose accuracy is traceable to EPA or NIST
standards.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...