Operator’s Manual
Auto Lock Time
Defines the number of seconds the pump will wait before automatically
locking the pump’s interface. Auto Lock Time is a numeric value in the range
of 0 to 999 seconds. Setting this value to zero (0) or
(none)
will disable the
Auto Lock
feature.
Bolus Change in mL
Defines if a pump user can change a
Bolus Dose
or
Loading Dose
in mL. Select
from two values-
Enabled
or
Disabled
.
Bolus Time Feature
Defines if a pump user will need to enter a time when setting a
Bolus Dose in
a manual delivery mode
.
Due to a conflict between Drug Program and Profile
options, bolus time is always enabled in Drug Programs
. Select from two
values-
Enabled
or
Disabled
.
Bolus Total Dose
Defines if a pump user will be allowed to enter a Bolus Dose or Loading Dose
as a Total Dose rather than the standard dose/kg in weight based modes. Select
from two values—
Enabled
and
Disabled
.
Date Separator
Defines whether dates are separated by a slash (yyyy/mm/dd) or a dash (yyyy-
mm-dd).
Decimal Separator
Defines whether decimal points in numbers are represented by a comma or a
period.
Enter User ID on Override
Defines if the pump will prompt a user for their
User ID
when overriding
a Drug Program infusion property. Select from two values—
Enabled
or
Disabled
.
Flush
Defines if the Flush features will be available on the pump. Select from two
values—
Enabled
or
disabled
.
Flush Time
Defines if a user will be prompted to enter a flush time when programming a
flush infusion. Select from two values—
Enabled
or
disabled
.
Maximum Flush Volume
Defines the maximum flush volume (in mL) allowed by the pump during a
flush infusion. Maximum flush volume is a numeric value from the minimum
volume for the syringe to the fill volume of the syringe.
FlowSentry™
Defines if the FlowSentry™ pressure monitoring (for rapid occlusion detection)
feature is enabled. Select from two values—
Enabled
or
Disabled
.
FlowSentry™ Sensitivity
Defines the sensitivity of the FlowSentry™ pressure monitoring algorithm
during normal infusion. Select from four values—
Low - least sensitive
,
Normal
,
High
, or
Very High - most sensitive
.
FlowSentry™ Startup Sensitivity
Defines the sensitivity of the FlowSentry™ pressure monitoring algorithm
during infusion startup. Select from four values—
Low - least sensitive
,
Normal
,
High
, or
Very High - most sensitive
.
Key Click (key click loudness)
Defines the click loudness level when pressing pump keys. Select from five
levels—
Off
,
Level 2
,
Level 3
,
Level 4
, or
Level 5
(Level 5 is the loudest).
Maximum Body Surface Area
Defines the maximum allowed value for the
Body Surface Area
infusion
parameter. Maximum Body Surface Area is a numeric value in the range of
0.01 to 4 m
2
.
Maximum Bolus Rate
Defines the
maximum
bolus infusion rate allowed by the pump. Select from
three rates—
Syringe Size x 5 mL/hr
(example: 5 mL syringe × 5 mL/hr = 25
mL/hr maximum bolus rate),
Syringe Size x 10 mL/hr
(example: 5 mL syringe