Medfusion® Model 3500 pump V6
Operator’s Manual
Override and toggle features
When the pump is paused at the
Begin Infusion
screen (or, in some cases while infusing), pressing
Options
provides access to several override and toggle
options that will affect the current infusion
only
.
These
Options
will return to the default selection
whenever the pump is turned off, or whenever the
program is returned to the
Main
menu prior to
beginning the next infusion.
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
BOLUS DOSE
5
OVERRIDE OCCL LIMIT
2
PERIODIC CALLBACK ALARM
6
CHANGE TO DOSE
3
OVERRIDE ALARM LOUDNESS
7
VOLUME LIMIT
4
DELAYED START
8
DISABLE FLOWSENTRY
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG DOSE
MAIN MENU
CONC
1.5
mg
/
m
L
PVD
0
m
L
WEIGHT
5
kg
DOSE
20
mg
TIME
02:00:00
TIME REMAINING
02:00:00
RATE
6.67
m
L/
Hr
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG DOSE
MAIN MENU
CONC
1.5
mg
/
m
L
PDD
0
mcg
WEIGHT
5
kg
DOSE
20
mg
TIME
02:00:00
TIME REMAINING
02:00:00
RATE
6.67
m
L/
Hr
PVD (Program Volume
Delivered) indicates
Volume
is
currently selected
PDD (Program Dose Delivered)
indicates
Dose
is currently
selected
Change to Dose / Change to Volume
When delivering a drug program or infusion
containing a concentration (must be a
“continuous” style infusion), you may want to view
either the volume delivered (PVD) or the dose
delivered (PDD). The toggle PVD / PDD option
may be used at the
Begin Infusion
screen, or
during an infusion without stopping medication
delivery.
1. With the pump at the
Begin Infusion
screen (or during medication delivery),
press
Options
.
2. Use the
number
keys to choose
Change
to Dose
or
Change to Volume
(only
one or the other will be shown, depending
on the current setting). (You may have to
press
More
to see this option.)
3. PVD will change to PDD (or vice-versa,
depending on the current setting). See page
97 for more information on PVD/PDD.