55
Medfusion® Model 3500 pump V6
Operator’s Manual
OPTIONS
CLEAR TVD
PRESS < « > TO BEGIN INFUSION
MAIN MENU
VOLUME
60
m
L
TVD
0.000
m
L
TIME
00:55:00
TIME BETWEEN STARTS
05:00:00
TIME REMAINING
00:55:00
RATE
65.5
m
L/
Hr
6. Confirm the settings and start the infusion,
or use the three menu softkeys shown at
right to modify the settings.
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
7. When all delivery settings are confirmed as
correct, press
É
to begin infusion.
Main Menu
– returns to the
M
ain
M
enu
.
Options
– allows selection from a list of variables (e.g.,
standby, etc.).
Clear TVD
– clears total volume delivered.