98
Medfusion® Model 3500 pump V6
Operator’s Manual
Clearing total volume, program volume, or program dose delivered
Use the following steps to clear the total volume and
program volume (or dose) and reset the counter to
zero.
With the pump paused (press
á
, if necessary):
1. Press
Clear Totals
.
2. Confirmation is required to clear totals
delivered. Press
Yes
to clear total volume
delivered and program volume (or dose)
delivered. (Or press
No
to exit without
clearing.)
The program volume (or dose) delivered is cleared
automatically whenever the pump’s delivery mode
is changed or following the
Infusion Complete
alarm. The Total Volume Delivered is only cleared
when the softkey is pressed (i.e.,
Clear Totals
key).
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CONC
1.2
mg
/
m
L
PVD
0.000
m
L
WEIGHT
50
KG
DOSE
2.5
MCG/KG/MIN
RATE
3.5
m
L/
Hr
CHG DOSE
MAIN MENU
NO
CLEAR TOTAL VOLUME DELIVERED?
YES
With the pump delivering:
1. From any run screen, press
Clear TVD
.
2. Confirmation is required to clear the Total
Volume delivered. Press
Yes
to clear total
volume delivered only. (Or press
No
to exit
without clearing.)
The program volume (or dose) delivered is cleared
automatically whenever the pump’s delivery mode
is changed or following the
Infusion Complete
alarm. The Total Volume Delivered is only cleared
when the softkey is pressed (i.e.,
Clear Totals
key).
Recovering TVD Information
: The Total volume
delivered display can be restored after turning the
pump off by using the
Recall Last Settings
selection
from the
Main
menu. If TVD has been cleared,
then only the quantity infused since clearing will be
stored in memory. It may be good practice to make
note of TVD before clearing it.
LOCK
N
PVD
5.048
m
L
KVO V LIMIT
60
m
L
RATE
3.5
m
L/
Hr
B-D
60
mL
CHG RATE
CLEAR TVD
NO
CLEAR TOTAL VOLUME DELIVERED?
YES