68
Medfusion® Model 3500 pump V6
Operator’s Manual
5. Use the
number
keys to set delivery time
in minutes & seconds. The default value
for loading (i.e., bolus) delivery rate is
set in custom configuration. Maximum
& minimum time is calculated from the
maximum & minimum rate for the syringe
size currently in use (or maximum and
minimum can be set using PharmGuard
®
Toolbox 2).
Using PharmGuard
®
Toolbox 2, it is
possible to restrict loading time to minutes
only. If that is the case the screen will look
like this:
Using PharmGuard
®
Toolbox 2, it is also
possible to expand loading time to up to
12 hours. If that is the case the screen will
look like this:
6. After the loading dose is set and accepted,
the pump pauses. The Loading Dose
setting now appears on the
Begin
Infusion
screen. If this is a new infusion,
prior to connecting to the patient press
è
to initiate the priming sequence (page
58).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy
of
all
infusion values to the original order.
Programming the pump at a delivery rate other
than that prescribed will cause inappropriate
delivery, which can result in serious injury or
death.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
CONC
20
mcg
/
m
L
PVD
0
m
L
WEIGHT
40
KG
KVO
LOAD
5
MG/KG
DOSE
33.3
MG/KG/HR
RATE
66.6
m
L/
Hr
BACKSPACE
ENTER
ENTER LOADING TIME - PRESS ENTER TO CONTINUE
HIGH: 00:59:00
LOW: 00:03:00
CLEAR
HR : MIN : SEC
00:
03:
00
00:03:00 CURRENT SETTING
BACKSPACE
ENTER
ENTER LOADING TIME - PRESS ENTER TO CONTINUE
HIGH: 00:59:00
LOW: 00:03:00
CLEAR
HR : MIN : SEC
00:
03
:
00
00:03:00 CURRENT SETTING
BACKSPACE
ENTER
ENTER LOADING TIME - PRESS ENTER TO CONTINUE
HIGH: 12:00:00
LOW: 00:03:00
CLEAR
HR : MIN : SEC
10:00
:
00
10:00:00 CURRENT SETTING