83
Medfusion® Model 3500 pump V6
Operator’s Manual
Override occlusion limit
The pump’s programmed default occlusion setting
can be overridden. This allows setting a higher or
lower occlusion limit for a specific infusion.
Note:
When a 1mL syringe is entered into the
programming, the occlusion pressure setting is
fixed
at
very high
(50psi), meaning there is no
option of changing it.
1. Set the delivery mode parameters. When
the pump is paused at the
Begin Infusion
screen, press
Options
.
2. Use the
number
keys to choose
Override
Occl Limit
. (You may have to press
More
to see this option.)
3. The
highlight
shows the pump’s current
default setting. Use the
number
keys to
choose the desired occlusion setting, then
press
Enter
(see the table on page 93 for
more information about occlusion settings).
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
STANDBY
5
DELAYED START
2
VOLUME LIMIT
6
LOADING DOSE
3
OVERRIDE OCCL LIMIT
7
BOLUS DOSE
4
OVERRIDE ALARM LOUDNESS
8
KVO
SELECT OCCLUSION LIMIT - PRESS ENTER TO CONTINUE
ENTER
1
VERY LOW - MOST SENSITIVE
2
LOW
3
NORMAL
4
HIGH - LEAST SENSITIVE
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr