113
Medfusion® Model 3500 pump V6
Operator’s Manual
Poleclamp use
This section contains the instructions for using the
poleclamp options available for the pump.
The poleclamp anchor is attached to the pump
according to the
Instructions for Use
supplied with the
poleclamp assembly.
CAUTION:
This pump is designed to be used in
a
horizontal
position. If the pump is operated in
any other position (including vertical), there is an
increased potential for fluid leaking into the pump. If
you suspect fluid has leaked into a pump, remove it
from service until a trained biomedical technician can
test it.
Rotating/detachable
poleclamp (standard)
Smiths Medical offers two rotating poleclamps,
one with a 90-degree angle anchor bracket (catalog
number 22-1050-51), and one with a 100-degree
angle anchor bracket (catalog number 22-1051-51)
(older pumps may use poleclamp catalog number
3000RPC, which is no longer available).
The 100-degree anchor bracket attaches the pump
to a pole at a slight forward angle, making the
display and keypad easier to view when pumps are
attached to a pole in certain circumstances (for
instance, high on a pole).
Align the poleclamp for mounting on any
vertical
or
horizontal
pole, bed rail, or similar solid
structure small enough to fit within the clamp’s
jaws. The light weight of the pump makes this
convenient.
1. Open the clamp jaws wide enough to easily
slip around the pole.
2. Tighten the clamp
until it is securely
fastened
so the pump cannot be moved.
CAUTION:
Always verify the stability of the object
to which the pump is mounted (for instance, an IV
pole) before using the poleclamp. Failure to verify
the stability could cause the object to tip, with the
possibility of causing damage to the pump and
other equipment. If the poleclamp is not adequately
tightened, it could cause the pump to fall.
90-degree poleclamp
100-degree poleclamp