130
Medfusion® Model 3500 pump V6
Operator’s Manual
Model
(Manufacturer)
Size
Min. Rate
Max. Rate Minimum volume
to infuse
Minimum recommended rate
mL
mL/hr
mL/hr
mL
mL/hr
BB Perfusor® (B. Braun)
Luer Lock
20
0.1
481
0.0333
0.67
0.2
50
0.1
1042
0.0833
1.7
0.5
BB Omnifix™ (B. Braun)
Luer Lock
5
0.03
207
0.0083
0.17
0.05
10
0.05
338
0.0166
0.33
0.1
20
0.1
537
0.0333
0.67
0.2
50
0.1
1042
0.0833
1.7
0.5
*FlowSentry™ pressure monitoring is not available
with 1 mL and Terumo
®
syringes.
Syringe Size
Minimum Recommended
Bolus Volume
1mL
0.1mL
3mL
0.3mL
5mL
0.5mL
6mL
0.6mL
10mL
1mL
12mL
1mL
20mL
1mL
30mL
1mL
35mL
1mL
50mL
1mL
60mL
1mL
WARNING:
Use the smallest syringe size
necessary to deliver the fluid or medication. Using
a large syringe at very low rates (below Minimum
Recommended Rate for the syringe) may cause
improper pump operation, delayed occlusion sensing,
larger post occlusion bolus at higher occlusion limit
settings, delivery inaccuracies, or other potential
hazards.
Bolus Volume:
Delivering a bolus volume
less than the recommended bolus volume for the
syringe used may result in delivery inaccuracies. Use
an infusion set with the smallest diameter tubing
available that does not result in excessive back
pressure at the desired flow rate. Consider priming,
loading, bolus, and flush rates when selecting an
infusion set.
Per the US Food and Drug Administration, the 28
syringes available as the general configuration for
the Medfusion
®
Model 3500 pump V6 were tested
for stability with Morphine Sulfate, Dopamine
and Floxuridine for Injection, USP as medications
representative of drugs used for the routes of
administration specified for use with the Medfusion
®
Model 3500 pump V6. The testing demonstrated that
the pump had no adverse affect on the syringes and
the medications contained within.
Other medications may be suitable for use with the
syringes referenced for use with the Medfusion
®
Model 3500 pump V6. Before using different
medications, refer to the FDA approved labeling by
the drug manufacturer and syringe manufacturer,
FDA requirements and your facility’s established
protocols and procedures for determining the
suitability and compatibility of a desired medication
for use with the syringes listed for use with the
Medfusion
®
Model 3500 pump V6.
Prior to using a syringe for a route of administration,
ensure that the syringe used meets the requirements
for sterility and pyrogenicity for the desired route of
administration.