Using Quick Libraries
Smiths Medical ships the infusion pump with the
Medfusion
®
Standard Configuration that does not
include a Quick Library.
The profiles, categories and custom drug programs
available on a pump depend on the configuration
set up by the person responsible for custom
configuration. Quick Libraries is an option that
can be set-up in custom configuration. This option
designates a category in a profile as a Quick Library.
A profile with Quick Library enabled displays a list of
custom drug programs.
1
P
roPoFoL
10
mg
/
m
L
2
v
ecUroniom
1
mg
/
m
L (or)
3
dobUt
amine
200
mcg
/
m
L (or)
4
doPa
mine
1600
mcg
/
m
L (or)
MAIN MENU
PRESS NUMBER TO SELECT
1/1
Rather than the
CHG Profile
key displayed on
standard library category screens, the Quick Library
will have
Main Menu
on the far left. When
Main
Menu
is pressed, another selection screen is displayed,
with the additional categories and delivery modes for
the selected profile.
MORE
1
Anesthesia A-D
5
Anesthesia Q-Z
2
Anesthesia E-I
3
Anesthesia J-M
4
Anesthesia N-P
CHG PROFILE
PRESS THE NUMBER TO SELECT
aneStHeSia
QUICK LIBRARY
The softkey on the far left of this screen reflects the
name of the Quick Library (the above is an example
only). Press this key to re-enter the Quick Library.
The Quick Library is exited when
Main Menu
is
pressed. The
CHG Profile
key responds the same here
as with all others.
(If there is only one profile on the pump, and a
Quick Library is enabled, the drug programs in the
designated Quick Library will display on screen after
the pump is turned on and completes its self tests.)
The drug programs in a Quick Library have
been customized with default values for infusion
parameters that may include: syringe model and
size, dose, volume, weight, delivery time, etc. Setting
default values in the custom program allows the user
to proceed from the drug program selection screen
to the final confirmation screen (provided the loaded
syringe matches the default syringe in the drug
program).
The Quick Library program’s preset values are
loaded automatically into the pump
provided that
you have loaded a syringe of the manufacturer and
size programmed for this library entry
. Press
É
to begin the preset infusion after verifying infusion
settings and using the
Prime
feature. Some data fields
may be blank and require data entry to proceed.