50
Medfusion® Model 3500 pump V6
Operator’s Manual
BACKSPACE
ENTER
ENTER DELIVERY TIME
HIGH: 00:30:00
LOW: 00:03:00
CLEAR
:00
00:00
H R : M I N : S E C
4. The time setting screen appears. Press the
number
keys to set the time in hours and
minutes, then press
Enter
.
5. To confirm the setting, press
Yes
. To make
changes before confirming, press
No
. At
the
Begin Infusion
screen all values are
set and the pump pauses (the pump may
be configured with additional confirmation
screens, requiring confirmation of all
settings again before beginning the
infusion).
6. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
7. Confirm the settings and start the infusion,
or use the menu softkeys shown at right to
modify the settings.
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
8. When all delivery settings are confirmed as
correct, press
É
to begin infusion.
MAIN MENU
CONC
100
mg
/
m
L
TVD
0
m
L
WEIGHT
5
KG
DOSE
250
MG
TIME
00:15:00
TIME REMAINING
00:15:00
RATE
10
m
L/
Hr
OPTIONS
CLEAR TVD
a
mPiciLLin
NICU Intermittent
Main Menu
– returns to the
M
ain
M
enu
.
Options
– allows selection from a list of variables (e.g.,
bolus, volume limit. etc.).
Clear TVD
– clears total volume delivered.
NO
CONFIRM SETTINGS?
YES
conc
100
mg
/
m
L
WeigHt
5
kg
doSe 250
mg
time 00:15:00
rate
1.5
m
L/
Hr