× 10 mL/hr = 50 mL/hr maximum bolus rate), or
Syringe Size Max mL/hr
(see
the table on page 129).
Maximum Flow Rate
Defines the maximum flow rate allowed by the pump. Maximum Flow Rate
is a numeric value in the range 0.1 to 1130 mL/hr, and is dependent on the
syringe size being used.
Maximum Patient Weight
Defines the maximum allowed value for the
Patient Weight
infusion
parameter. Maximum Patient Weight is a numeric value in the range 0.25 to
250 kg.
Near Empty Time
Defines the number of minutes remaining before a syringe is empty, at which
time a
Syringe Near Empty Alarm
will occur. Near Empty Alarm Time is a
numeric value in the range 0 to 240 minutes. Setting this value to zero will
disable the
Syringe Near Empty Alarm
.
Near Empty Alarm Priority
Defines the priority level at which to generate the
Syringe Near Empty Alarm
.
Select from two levels—
Low
or
Medium
.
Occlusion Limit
Defines the pressure at which the pump will consider the line occluded. Select
from four limits—
Very Low (Approx. 4 psi, 206.9 mmHg)
,
Low (Approx. 8 psi,
413.7 mmHg)
,
Normal (Approx. 12 psi, 620.6 mmHg)
, and
High (Approx. 16
psi, 827.4 mmHg)
. For 1 mL syringes, the occlusion limit is fixed at
Very High
(Approx. 50 psi) and cannot be changed.
Profile Passcode
Allows setting a
four-digit passcode
which the user must enter to access the
Profile. Profile Passcodes can be any numeric code in the range of 1000 to
9999. If set to
(none)
, no passcode is required to access the Profile.
Program Display
Defines if the pump will display the delivered
Program Dose
or the delivered
Program Volume
. Select from two values:
Volume
or
Dose
.
Quick Library
Defines the Library Category that will be used as the Quick Library.
Syringe 1mL Support
Defines if 1 mL syringe support is enabled, allowing the use of 1 mL syringes.
Select from two values—
Enabled
or
Disabled
.
Unlock Passcode
Defines the passcode that must be entered at a pump to unlock its interface.
Unlock Passcodes can be any four-digit numeric code in the range 1000 to
9999. If set to
(none)
, no passcode is required to unlock the pump.
Auto Prompt Options
Auto Prompt Options are optional settings that the
user will be prompted to enter for all infusions initiated
in the defined profile. The default for all Auto Prompt
Options is Disabled.
Bolus Dose
Defines if a pump user will be prompted to enter a
Bolus Dose
during infusion
setup. Select from two values—
Enabled
or
Disabled
.
Confirm Settings
Defines if a pump user will be prompted to
Confirm Infusion Settings
when
setting up an infusion. Once the infusion settings are made, if Confirm
Settings is enabled the user will be required to re-confirm the settings. Select
from two values—
Enabled
or
Disabled
.
Enter User ID to Program
Defines if a pump user will be prompted to enter their
User ID
when
programming an infusion. Select from two values—
Enabled
or
Disabled
.