25
Medfusion® Model 3500 pump V6
Operator’s Manual
Programming an Infusion
Turn the pump on
Below right is an illustration of a startup screen, and a
list of basic steps for turning the infusion pump on.
1.
Press and hold
ë
until the display turns
dark. Verify the display shows all pixels on,
then all pixels off.
2.
The pump “beeps” twice, one low beep
and one high beep. This is a test of the
standard audible alarm and the battery
backup audible alarm (high tone). The
Alarm, Infusing and Lock indicators
flash as the self-test cycles and the startup
screen appears. The Device ID, if it has
one, appears in the middle left corner. (In
this example, it is called “HOSPITAL”.)
The pump serial number will appear at
the middle right (in this example, it is
“
M001234
”).
3.
Verify the display indicates successful
completion of system startup self-tests.
If the pump does not turn on with successful
completion of self-tests
, remove it from service
and have it checked by a trained biomedical
technician.
IMPORTANT:
Do not move the Syringe Plunger
Driver or otherwise manipulate the pump until
the startup tests are complete and the Profile
selection screen appears. Sensors are tested as part
of the startup process, and manipulating the pump
during this process may cause it to detect false
sensor failures, resulting in alarms.
VERIFYING CONFIGURATION
P
HARM
G
UARD
®
M
EDICATION
S
AFETY
HOSPITAL
VERSION
V6.0.0
M001234
CONFIG VERSION:
01.05.00002
CONFIG CRC: 4A4F
5
SECONDS REMAINING
©
99-2013
100%