Operator’s Manual
System Failure Alarms
Alarm Message
Priority
Remedy
System Failure :
Control Key Switch BGND
Test
System
Fault
Either a key on the keypad has been pressed for longer than 7 seconds,
or the keypad needs to be replaced; fluid delivery stops. Press
ë
to
turn the pump off, then turn it back on. If the alarm persists when the
pump is turned back on,
immediately remove pump from service
for
repair by a trained biomedical technician.
System Failure:
Configuration Required
System
Fault
The pump has detected that a Configuration is not loaded onto the
pump, and the pump cannot be used. A Configuration must be
loaded onto the pump either using PharmGuard
®
Toolbox 2 or the
Configuration Cloning procedure. Remove the pump from service and
replace with another pump which has a validated Configuration.
System Failure:
Force Sensor Test
System
Fault
A self-test was performed on the force sensor, and the test failed due to
a sensor malfunction
OR
the Syringe Plunger Driver was being moved
during the pump startup process, or the Syringe Plunger Driver received
a hard bump during the pump startup process. Press
í
. Press
ë
to turn the pump off, then turn the pump back on. If alarm persists,
remove the pump from service for repair by a trained biomedical service
technician.
System Failure:
Supercap Post
System
Fault
Power-up self-test sensed insufficient charge in the super-capacitor.
This problem can occur when a new Main PCB is installed or when a
pump battery is totally dead. Plug the pump into AC power for at least
two minutes then turn the pump off, then back on. If the alarm persists
when the pump is turned back on,
immediately remove pump from
service
for repair by a trained biomedical technician.
System Failure :
xx xx xx
System
Fault
Where “
xx xx xx
” indicates additional text for the specific type of failure.
The pump display shows the
Main
menu, with only the
Biomed
menu available. Some system failures result in a hardware “watchdog”
alarm which results in an alarm tone sounding (and may or may not
be accompanied by a flashing red alarm indicator). Press
ë
to turn
off pump, then back on. If the alarm persists when the pump is turned
back on,
immediately remove pump from service
for repair by a trained
biomedical technician.
System Advisory Alarms
Alarm Message
Priority
Remedy
System Advisory :
Improper Shutdown
Low
The pump improperly powered down. As soon as possible remove pump
from service for repair by a trained biomedical technician.
System Advisory :
Maintenance is
Recommended
Low
The scheduled Preventive Maintenance is due. As soon as possible,
remove pump from service for testing and checkout by a trained
biomedical technician. This alarm will reactivate each time the pump is
powered on until service is performed.
System Advisory :
Set Time and Date
Low
The pump’s Time and Date settings have been lost. As soon as possible,
remove pump from service for testing and checkout by a trained
biomedical technician.