38
Medfusion® Model 3500 pump V6
Operator’s Manual
Programming Infusions
Continuous mode: mL/hr
Continuous infusion is programmed by setting a
flow
rate in mL/hr
. Actual range limits for flow rate are
determined by syringe model and size (see the table
in Technical section (page 129) of this manual and/
or the product literature supplied with PharmGuard
®
Toolbox 2 or the PharmGuard
®
Supported Syringes
[PSS] Series 2) or by a custom configured maximum
rate. If needed, and if the options have been enabled
using PharmGuard
®
Toolbox 2, a volume limit,
loading dose, or bolus dose can also be programmed.
To program a Continuous mode infusion, at the
selection menu, choose
ml/hr
(or choose a drug
program that uses the continuous mode infusion
type). Choose the syringe model and size, and load
the syringe as instructed on page 33.
1. Use the
number
keys and (if needed) the
decimal point to set the flow rate. Press
Clear
to clear an entry. Press
Backspace
to
erase a character. Press
Enter
to accept the
setting.
2. After all infusion values are set, the pump
pauses at the
Begin Infusion
screen
(the pump may be configured with
additional confirmation screens, requiring
confirmation of all settings twice before
beginning the infusion).
3. If desired, program a Bolus Dose, Loading
Dose or Volume Limit (see Options, page
60).
4. Prior to connecting to the patient, press
è
to initiate the priming sequence (page 58).
5. Confirm the settings and start the infusion,
or use the four menu softkeys to modify the
settings.
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
6. When all delivery settings are confirmed as
correct, press
É
to begin the infusion.
BACKSPACE
ENTER
ENTER RATE - PRESS ENTER TO CONTINUE
HIGH: 944
LOW: 0.1
CLEAR
m
L /
Hr
50
OPTIONS
CLEAR TOTALS
PRESS < « > TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0
m
L
RATE
50
m
L/
Hr
Main Menu –
returns to the
M
ain
M
enu
.
Chg Rate –
allows modifying the set rate.
Options –
selection from a list of variables (e.g., bolus,
volume limit, etc.).
Clear Totals
–
clears total volume delivered, and PVD or
PDD.