54
Medfusion® Model 3500 pump V6
Operator’s Manual
Intermittent volume / time
Intermittent Volume / Time
mode allows specification
of a
delivery volume
for a
specific delivery time
, after
which the delivery
stops
for a programmed interval,
then the pattern recycles. In other words, the pump
delivers for a set period, pauses a set period, then
infuses for a set period, and so on.
• The
delivery stop interval
equals “
Time Between
Starts
” minus “
Delivery Time
.”
• The pattern
repeats
at the “Time Between Starts”
interval.
Below are the steps for programming an Intermittent
Volume / Time infusion. Select
Intermittent
Volume / Time
from the selection menu (or choose
a drug program that uses the intermittent volume/
time mode infusion type). Choose the syringe model
and size, and load the syringe as instructed on page
33.
1. Use the
number
keys to set the volume.
Volume is limited to 500 mL, and minimum
volume is 1/600
th
of syringe size. When the
volume is correct, press
Enter
.
Note:
Volume is limited to 500 mL regardless of
syringe size in use, as some infusions may require
multiple syringes.
2. Use the
number
keys to set the delivery
time. Maximum and minimum times are
calculated for the syringe in use and/or the
custom setup in the Configuration. When
ready, press
Enter
.
3. Use the
number
keys to set the time
between starts value. Maximum and
minimum times are calculated for the
delivery time entered (minimum time
must be at least one minute longer than
the delivery time; the pump will not accept
time between starts equal to or less than the
delivery time). When ready, press
Enter
.
4. At the
Begin Infusion
screen all values
are set and the pump pauses (the pump may
be configured with additional confirmation
screens, requiring confirmation of all
settings again before beginning the
infusion).
5. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
BACKSPACE
ENTER
ENTER VOLUME TO INFUSE - PRESS ENTER TO CONTINUE
HIGH: 500
LOW: 0.1
CLEAR
m
L
60
BACKSPACE
ENTER
ENTER DELIVERY TIME
HIGH: 24:00:00
LOW: 00:04:00
CLEAR
:00
00:55
H R : M I N : S E C
BACKSPACE
ENTER
ENTER TIME BETWEEN STARTS
HIGH: 24:00:00
LOW: 00:06:00
CLEAR
:00
05:00
HR:MIN:SEC