Alarm Message
Priority
Remedy
Motor Not Running
High
The motor is not running, which may be due to an erroneous (possibly
transient) sensor signal, or a mechanical and/or electrical failure. Press
í
. Press
ë
to turn the pump off, then back on. If the alarm persists
when the pump is turned back on,
immediately remove the pump from
service
for repair by a trained biomedical service technician.
Motor Rate Error
High
The motor is not running at the programmed rate, which may be due
to an erroneous (possibly transient) sensor signal, or a mechanical and/
or electrical failure. Press
í
. Press
ë
to turn the pump off, then back
on. If the alarm persists when the pump is turned back on,
immediately
remove the pump from service
for repair by a trained biomedical service
technician.
Occlusion - Check Infusion
Line
High
Check infusion line for blockage. Clear or remedy any discovered
occlusion; for example, check the infusion set for closed stopcocks,
clamps, check the tubing for kinks, verify the patency of the tubing
and the IV site, etc. Press
É
to restart infusion once the occlusion is
located and cleared. If this alarm occurs frequently and you have ruled
out an actual occlusion (closed clamp, etc.), it may be the infusion set
tubing is too small for the rate or fluid viscosity, and you may want
to consider enabling FlowSentry™ pressure monitoring, or choosing a
different occlusion alarm setting in the Configuration. If you cannot
clear the occlusion alarm, there may be a problem with the pump that
requires repair by a trained biomedical technician.
Outside Range Limit -
Silence Alarm to Continue
Medium
The entered rate is outside of the Drug Program soft limits. Press
í
to
continue. You may then choose to override the range limit or change the
value entered
.
Periodic Callback - Silence
Alarm to Continue
Medium
Press
í
. User programmed callback alarm - no remedy required.
PharmGuard Data Transfer
is Recommended
Medium
As soon as possible, remove pump from service and transfer
PharmGuard
®
Event Data to a host computer. Continued use may result
in lost PharmGuard
®
Event Data.
Pressure Increasing -
Check Infusion Line
Medium
Press
í
. FlowSentry™ pressure monitoring detects increasing pressure
in the infusion line. Check infusion line for blockage; for example, check
the infusion set for closed stopcocks, clamps, check the tubing for kinks,
verify the patency of the tubing and the IV site, etc. Clear or remedy any
discovered occlusion. If occlusion is not cleared, the “
Occlusion
” alarm
will result.
Profile Does Not Match
Last Settings Profile
Limit
The user attempted to use the “Recall Last Settings” feature. Recall Last
Settings only works if the pump is in the same profile under which the
previous infusion was programmed.
Pump is Locked
Limit
The keypad has been locked to prevent changes during infusion. The
alarm is initiated by pressing any key except
Unlock
. Press
Unlock
to
make changes to delivery settings.
Rate Below
Recommended Min for
Syringe Size
Limit
The programmed rate is below that recommended for the syringe size
- no remedy required (see
Guidelines for enhanced pump performance
,
page 23).