53
Medfusion® Model 3500 pump V6
Operator’s Manual
Volume / time
The volume over time rate calculation allows
programming by:
•
Dose Volume
•
Delivery Time
The pump calculates the correct flow rate by dividing
the volume by the delivery time. Volume limit,
loading or bolus dose
cannot
be set in this mode.
To program Volume / Time, choose it at the selection
menu (or choose a drug program that uses the
volume/time mode infusion type). Choose the syringe
model and size, and load the syringe as instructed on
page 33.
1. Press the
number
keys to program the
volume to infuse. Volume is limited to 500
mL – and minimum volume is 1/600
th
of
the selected syringe size. When the Volume
setting is correct, press
Enter
.
Note:
Volume is limited to 500 mL regardless of
syringe size in use, as some infusions may require
multiple syringes.
2. Press the
number
keys to set delivery
time. Time is always limited by maximum
and minimum rates for selected syringe
size and/or the custom setup in the
Configuration. When the delivery time is
correct, press
Enter
.
3. At the
Begin Infusion
screen all values
are set and the pump pauses (the pump may
be configured with additional confirmation
screens, requiring confirmation of all
settings again before beginning the
infusion).
4. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
5. Confirm the settings and start the infusion,
or use the three menu softkeys to modify
the settings.
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
6. When all delivery settings are confirmed as
correct, press
É
to begin infusion.
BACKSPACE
ENTER
ENTER VOLUME TO INFUSE - PRESS ENTER TO CONTINUE
HIGH: 500
LOW: 0.005
CLEAR
m
L
50
BACKSPACE
ENTER
ENTER DELIVERY TIME
HIGH: 24:00:00
LOW: 00:31:00
CLEAR
:00
00:40
H R : M I N : S E C
OPTIONS
CLEAR TVD
PRESS < « > TO BEGIN INFUSION
MAIN MENU
VOLUME
50
m
L
TVD
0
m
L
TIME
00:40:00
TIME REMAINING
00:40:00
RATE
75
mL/hr
Main Menu
– returns to the
M
ain
M
enu
.
Options
– allows selection from a list of variables (e.g.,
standby, etc.).
Clear TVD
– clears total volume delivered.