7. Squeeze the Plunger Release Lever on the
syringe plunger driver and push it toward
the syringe plunger; once it is flush with
the syringe plunger release the lever. Be
sure both holders close around the syringe
plunger.
8. Turn and lower the Barrel Clamp onto the
barrel of the syringe.
WARNINGS:
• Verify that the plunger holders securely capture
the syringe plunger. Make sure to capture the
syringe barrel and flange. Failure to properly
secure the syringe could result in uncontrolled
fluid flow to the patient, which could result in
serious injury or death.
• Always use the
Prime
function on the pump
when loading a new syringe in order to remove
any mechanical slack. Failure to prime correctly
can delay infusion delivery and cause Total
Volume Delivered to read higher than the
amount actually delivered to the patient.
The size should now appear in the display.
Note:
If the syringe is incorrectly loaded, guide
arrows appear on screen to identify the problem.
CONFIRM
CONFIRM SYRINGE MODEL / SIZE/ LOADING
B-D
60
m
L
CONFIRM
CONFIRM SYRINGE MODEL / SIZE/ LOADING
B-D
¤
≤
™
Arrow pointing down
“
™
“ means check syringe barrel clamp.
Arrow pointing left
“
≤
“ means check syringe plunger holders.
Arrow pointing up
“
¤
” means check syringe flange clip.