66
Medfusion® Model 3500 pump V6
Operator’s Manual
past the maximum bolus time, the bolus is
cancelled
.
• Press
í
to silence the alarm.
The
normal infusion delivery is continued
.
Determine the reason the bolus was
unable to be delivered before re-
attempting a bolus. Once the reason
for the bolus cancellation is cleared,
bolus delivery may be re-initiated (if
desired). Unless the bolus settings are
reprogrammed (either by entering the
Options
menu and reprogramming
the bolus Dose and/or Time, or by
pressing the
CHG Bolus
key at the
Begin Bolus Delivery?
screen and
reprogramming the bolus Dose and/or
Time), the last bolus rate setting from
just prior to the bolus cancellation will
continue to be applied. Reprogram the
bolus, if desired.
OPTIONS
CLEAR TOTALS
RESTRICTED FLOW - BOLUS CANCELLED
CHG DOSE
MAIN MENU
CONC
10
mg
/
m
L
PVD
12.763
m
L
WEIGHT
40
kg
KVO MAX
9
m
L/
Hr
boLUS
2
mg/mL
DOSE
4
mg
/
kg
/
Hr
RATE
16
m
L/
Hr
CHG BOLUS
NO
BEGIN BOLUS DELIVERY?
BOLUS
2
MG/KG
VOLUME
8
m
L
TIME
29:54
MM:SS
RATE
16.1
m
L/
Hr
YES
totaL
boLUS
doSe
80
mg