56
Medfusion® Model 3500 pump V6
Operator’s Manual
Recall last settings
The Medfusion
®
Model 3500 pump V6 will recall the
settings used to run the last infusion.
Note:
If a pump has a custom configuration (created
using PharmGuard
®
Toolbox 2 and taught to the
pump) the
Recall last settings
feature is available
only in those profiles it has been included in.
To use
the feature, first enter the profile for the previous
infusion
. Any optional Volume Limit, KVO, Bolus
Dose or Loading Dose settings are retained in the
infusion when this feature is used. All other options
(Standby, Override Occl Limit, etc.) are discarded and
must be reprogrammed from the
Options
menu at
the
Begin Infusion
screen.
Reuse the same delivery settings to repeat an identical
dosage of medication or fluids to a patient (choose
the syringe model and size, and load the syringe as
instructed on page 33):
1. Select
Recall Last Settings
from the
selection menu. (If it isn’t shown, press
the
More
key and search for it.) In this
example, press the number 4.
When selected, press
Enter
to confirm
the syringe type and size if the syringe has
been replaced. After the syringe has been
confirmed the
Continue Same Infusion?
screen appears.
2. Prior to connecting to the patient, press
è
to initiate the priming sequence (page
58).
3. If
No
is pressed, the pump returns to the
Main
menu to select a different Drug
Program or manual delivery mode.
If
Yes
is pressed, the pump pauses at the
Begin Infusion
screen and PVD is
not
cleared. After reviewing these previously
used settings, the pump pauses. To clear the
TVD and PVD/PDD, press
Clear Totals
.
(For volume limited infusions, only
the remaining volume specified will be
delivered.)
MORE
PRESS THE NUMBER TO SELECT
1
m
L /
Hr
2
VOLUME / TIME
3
DOSE / KG/ MIN
4
RECALL LAST SETTINGS
MAIN MENU
NO
CONTINUE SAME INFUSION?
YES
VOLUME
10
m
L
PVD
0.487
m
L
WEIGHT
75
kg
KVO
LOAD
2000
mcg
/
kg
DOSE
75
mcg
/
kg
/
min
RATE
33.8
m
L/
Hr