Operator’s Manual
Disable/enable vol empty alarm tone
The pump may be configured to have an alarm
sound to indicate when the volume of medication
in a syringe is near empty. (The calculation for
when
the alarm sounds is based on the syringe size
used and the volume initially programmed.) Use
this option to enable or disable this feature for this
infusion.
1. With the pump running or stopped, press
Options
.
2. Use the
number
keys to choose
Disable
Vol Empty Tone
or
Enable Vol Empty
Tone
(only one of the above will be
displayed, depending on whether the Empty
Alarm feature is currently active or not).
(You may have to press
More
to see this
option.)
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
BEGINNING
1
DISABLE F
LOW
S
ENTRY
2
DISABLE NEAR EMPTY TONE
3
DISABLE VOL EMPTY TONE
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr