80
Medfusion® Model 3500 pump V6
Operator’s Manual
Periodic callback alarm
The Periodic callback alarm is an
Option
that causes
an alarm to be generated at a programmed period of
time.
Note:
During delivery of a loading dose, bolus dose,
or flush delivery, the periodic callback timer is
suspended, then restarts once the optional delivery is
complete. For example, if a periodic callback is set for
1 hour, then a bolus dose is delivered which lasts 10
minutes, the periodic callback will actually occur at 1
hour and 10 minutes.
Programming periodic callback
alarm
1. With the pump delivering an infusion or
paused, press
Options
.
2. Use the
number
keys to choose
Periodic
Callback Alarm
. (You may need to
press
More
to see it.)
3. Set the callback alarm time in hours &
minutes. The maximum callback time is 8
hours. The default callback time is 2:00 Hr.
When the callback time is correct, press
Enter
to accept the setting and go to the
Begin Infusion
screen.
4. If this is a new infusion, prior to connecting
to the patient press
è
to initiate the
priming sequence (page 58).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
BOLUS
2.5
m
L
V LIMIT
50
m
L
KVO RATE
0.6
m
L/
Hr
RATE
3.5
m
L/
Hr
PRESS THE NUMBER TO SELECT THE PROGRAM OPTION
MORE
1
DISABE F
LOW
S
ENTRY
5
CHANGE TO DOSE
2
DISABLE NEAR EMPTY TONE
6
DISABLE FLOWSENTRY
3
DISABLE VOL EMPTY TONE
4
PERIODIC CALLBACK ALARM
BACKSPACE
ENTER
ENTER PERIODIC CALLBACK ALARM TIME - PRESS ENTER
HIGH: 08:00:00
LOW: 00:00:00
CLEAR
HR : MIN : SEC
02:00
:00
Enable / Disable Periodic Callback Alarm
: The
Periodic Callback Alarm
option can be enabled
/ disabled through custom configuration (using
PharmGuard
®
Toolbox 2).