91
Medfusion® Model 3500 pump V6
Operator’s Manual
Volume and time
1. When the programmed infusion is
complete or the
Syringe Empty
alarm
activates and the pump alarms, press
í
.
2. The
Begin Flush Setup
screen appears.
If flush is not desired at this point, press
No
(the pump will return to the
Begin
Infusion
screen). To begin the flush setup,
press
Yes
.
3. Remove the syringe at the end of the
infusion when prompted and replace with
the flush syringe. Use the
number
key to
select the syringe model.
4. Load the syringe onto the pump, verify the
correct size is shown, then press
Enter
.
5. Use the
number
keys to enter the desired
amount of flushing infusate, then press
Enter
.
6. The pump requires confirmation that
the flush be delivered at the same rate as
the previous infusion. To deliver at the
same rate, press
Yes
(the pump will then
automatically calculate the delivery time).
To deliver at a different rate, press
No
.
7. If
No
was pressed, the
Flush Time
screen
appears. Use the
number
keys to select the
delivery time, then press
Enter
.
OPTIONS
CLEAR TVD
INFUSION COMPLETE
PVD
1
m
L
WEIGHT
1
KG
TIME
00:03:00
VTBI
1
m
L
TIME REMAINING
00:00:00
RATE
20
m
L/
Hr
MAIN MENU
CHG DOSE
NO
BEGIN FLUSH SETUP?
YES
SELECT SYRINGE MODEL
1
B-D
2
MONOJECT
3
TERUMO
ENTER
LOAD SYRINGE - PRESS ENTER WHEN READY
B-D
5
m
L
BACKSPACE
ENTER
FLUSH VOLUME TO INFUSE - PRESS ENTER TO CONTINUE
HIGH: 3
LOW: 0.005
CLEAR
m
L
NO
CONFIRM SAME RATE?
YES
BACKSPACE
ENTER
FLUSH TIME - PRESS ENTER TO CONTINUE
HIGH: 00:10:00
LOW: 00:02:00
CLEAR
H R : M I N : S E C
00:00
:00