49
Medfusion® Model 3500 pump V6
Operator’s Manual
Dose/kg/time (dose/m
2
/time)
The body weight/surface area infusion delivery mode
will only be available on a pump
through custom
programming of a drug program via PharmGuard
®
Toolbox 2 (in other words, ‘Dose/Kg/Time’ and
‘Dose/m
2
/Time’ are never available as menu items,
they must be utilized in the creation of a drug
program, which is then sent to the pump). Once
created using PharmGuard
®
Toolbox 2, taught to a
Medfusion
®
Model 3500 pump V6, and once the drug
program has been selected from the library, entries on
the pump may be required for the following:
•
Drug selection only
•
Drug concentration (NOTE:
If preset limits
were programmed in PharmGuard
®
Toolbox 2,
this option will be available. If
a preset value
was
programmed in PharmGuard
®
Toolbox 2, the
concentration limits are not needed.
)
•
Patient’s weight/Body surface area
•
Dose
•
Delivery time
The pump calculates the correct flow rate in mL/hr to
achieve the required drug dose. The patient weight/
body surface area is used to calculate dose limits. To
program a body weight/surface area infusion, choose
a Drug Program with
Dose/Kg/Time
(
Dose/m
2
/
Time
). Choose the syringe model and size, and load
the syringe as instructed on page 33.
The following procedure is an example of how an
infusion might be programmed. The actual procedure
depends on how the infusion was initially customized.
1. Press the
number
keys to set medication
concentration. Concentration is limited to
4 digits. Press
Clear
to clear an entry. Press
Backspace
to erase a character. Press
Enter
to accept the setting.
2. Press the
number
keys to set the patient’s
weight in kilograms or surface area in
square meters. Weight/surface area are
limited to the indicated range. Press
Enter
to accept the patient weight/surface area
setting.
The pump may be customized with an
additional patient weight/body surface
area confirmation screen. If so, re-enter the
patient weight/body surface area, then press
Enter
to accept the setting.
3. At the
Dose
screen, press the
number
keys
to set the dose. Dose limits displayed are
dependant on custom settings and installed
syringe size. When the setting is correct
press
Enter
to accept the dose. (The pump
calculates the correct flow rate in mL/hr to
achieve the required drug dose.)
BACKSPACE
ENTER
ENTER CONCENTRATION - PRESS ENTER TO CONTINUE
HIGH: 250
LOW: 100
CLEAR
mg
/
m
L
BACKSPACE
ENTER
ENTER PATIENT WEIGHT
-
PRESS ENTER TO CONTINUE
HIGH: 10
LOW: 0.25
CLEAR
KG
BACKSPACE
ENTER
ENTER DOSE - PRESS ENTER TO CONTINUE
HIGH: 500
LOW: 100
CLEAR
MG
BACKSPACE
ENTER
RE-ENTER PATIENT WEIGHT - PRESS ENTER TO CONTINUE
HIGH: 250
LOW: 0.25
CLEAR
KG