74
Medfusion® Model 3500 pump V6
Operator’s Manual
4. The volume limit setting now appears on the
display. Program Volume Delivered (PVD)
will appear on the screen directly above the
V Limit
.
5. If this is a new infusion, prior to connecting
to the patient press
è
to initiate the
priming sequence (page 58).
WARNING:
Confirm All Settings
. Before starting
any
delivery,
always
confirm the accuracy of
all
infusion values to the original order. Programming
the pump at a delivery rate other than that
prescribed will cause inappropriate delivery, which
can result in serious injury or death.
6. Press
É
to begin infusion. The infusion
will automatically stop when it reaches the
preset volume limit.
OPTIONS
CLEAR TOTALS
PRESS < « > KEY TO BEGIN INFUSION
CHG RATE
MAIN MENU
PVD
0.000
m
L
V LIMIT
50
m
L
RATE
3.5
m
L/
Hr