Medfusion® Model 3500 pump V6
Operator’s Manual
delivery of drug are performed redundantly.
The two calculated values are then compared. If the
two values do not match, the software will display a
system fault screen, turn on a continuous two-tone
audible alarm, and stop all drug delivery.
Timer Data Registers - Time base BGND, time of
day clock POST, time of day clock timeout
The data in the Real Time Clock is checked at regular
intervals. If the data is not reasonable, the software
will turn on a continuous two-tone audible alarm and
stop all drug delivery.