REA
VERIFIER
© REA Elektronik GmbH, 64367 Mühltal, Germany --- REA MLV-2D Manual Version 3 – 21/12/2015
Page 60 of 98
Indicates how many data modules (single matrix cells) could be corrected by error
correction. Since error correction relates to code words and not individual matrix cells
these values have a large range (depending on distribution of errors).
Total number of pad characters (correct/erroneous): 2 (2, 0)
Specifies how many pad characters (i.e. used to pad unused code words) have been
deployed to pad out the code matrix.
9.9
Error list
Whenever errors are detected during code verification or warnings have been issued
the right-hand side list with errors and warnings appear.
Each of the error messages
listed itself leads to a
downgrading of the code
grade. Without resolving
these errors, no better grade
will be obtainable for the
code.
The messages included in the
"Warnings" group do not lead
to a downgrading. Warnings
are related to non-compliance with optional parameters, which are displayed for
information only but are not directly included in the grading of the code's quality.
Such items of information can be very useful for optimising code quality.
9.10
Verification report
The verification report can be printed as a graphical
verification report by using the File -> Print screen
function. The printout is a simulated screen dump,
enhanced by including the device parameters as header
text. The normal print function generates the verification
report in its long or short form (preset: Configure ->
Options -> Print). All formats can be used with a PDF
writer to generate a PDF file or printed out on an office
printer.