Device properties
This setting allows a basic choice between whether or
not to use the 90° coaxial lighting. The 90° coaxial
lighting is a paid hardware option. The 90° coaxial
lighting cannot be retrofitted but must be ordered
together with a new device. Detailed settings for the
lighting must be configured using device configuration
in TransWin32.
USB update
If an update has been downloaded (see section 16), the download also includes the
firmware for the MLV-2D. The directory "Device_Update" is contained within the
compressed, downloaded archive (ZIP file). This contains the update files (e.g.
update.rvu). Depending on the software version, variations may occur in file names
and numbers.
For the USB update, all files must be copied from this directory onto a USB pen drive.
Subdirectories must not be used. The USB pen drive does not need to be empty. Once
the function has been selected, "Application Update" and "Language Update" appear
for selection. Here, "Application Update" must always be selected. In special cases,
program text (only) can be corrected/amended by using "Language Update".