APPLIED OPERATIONS
/Protocol processor
OPERATION INSTRUCTIONS
2 - 1 7 5
Page
2-4-6 Loading a protocol of previous format
FLUOVIEW software Ver4.2 inform the protocol data conversion when the software reads
previous version of the data.
The rule of the conversion is explained below.
Rule A
•
Before Ver4.2, Frame = XYT or XYZT is converted to XYT or XYZT in count of Z slice
parameter.
•
If Z Slices = 1, the Frame becomes XYT.
•
If Z Slices is more than 1, the Frame becomes XYZT.
•
If the Frame is continuous one, Frame becomes Series.
Fig. 2-50 Sample Program (FRAP)
Rule B
•
Before Ver4.2, Frame (Separate) is converted with condition of Track and Z slices.
•
If Condition = 1, and Z slice =1, Frame becomes XY.
•
If Condition is more than 1, and Z slice = 1, Frame becomes XYT.
•
If Condition = 1, and Z slice is more than 1, Frame becomes XYZ.
•
If Condition and Z slice are more than 1, Frame becomes XYZT.