APPLIED OPERATIONS
/Image Acquisition
OPERATION INSTRUCTIONS
2 - 2 8
Page
10 Setting a Lower Scan Speed
1. The scan speed can be decreased using the scale in the [Scan Speed] group box on
the [Acquire] panel.
In general, setting a lower scan speed allows the acquired image quality to be
improved.
However, a low scan speed also lengthens the time required for image
acquisition.
When the scan speed is decreased during fluorescence observation, the
saturation of fluorescence may darken the image of certain types of
specimens. In this case, increase the scan speed and increase the PMT
Voltage or use accumulation in scanning.
11 Stopping Repeated Scanning
1. After the brightness and gain have been adjusted, select the <STOP SCAN> button in
the [Acquire] panel to stop scanning temporarily.
2-2-1-5 Acquiring Image
1. Select the <Once> button. The acquired image will be displayed in the [Live] panel.
<Once> button
TIP
NOTE
[Scan Speed] group box
Set the scan speed by clicking a point on the
scale line.