Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
OPERATION INSTRUCTIONS
1 - 3 0
Page
1-2-7-5 Setting the XY Observation Mode
1.
In the [Mode] group box in the [Acquire] panel, select the [Surface] option button.
2.
In the [Mode] group box in the [Acquire] panel, select [800 by 600] from the [Scan Size]
drop-down list.
3.
In the [Acquire] panel, select the XY observation mode option button.
1-2-7-6 Repeated Scanning Operation
1. Select the <XY Repeat> button. The acquired image will be displayed in the [Live]
panel.
Use the <FOCUS> button to acquire image at an even higher speed. If the
specimen is already being scanned, stop scanning with the <STOP
SCAN> button before selecting the <XY Repeat> button.
The Focus function reduces the scanning time by line skipped scan.
As a result, the acquired images become coarse.
Do not move FLOUVIEW FV1000 Menu while acquiring an image.
<XY Repeat> button
<Focus> button
NOTE
TIP
NOTE