APPLIED OPERATIONS
/Protocol processor
OPERATION INSTRUCTIONS
2 - 1 8 3
Page
z
Setting Track 3
In XYT observation, return the zoom ratio to 2X and acquire 30 images every 5 seconds.
(Observe the change in fluorescence image with time after bleaching.)
1.
Select “Series mode” in [Mode] of [Track].
2.
Select “Normal” in [Start by] of [Track].
3.
Enter “30” to [Condition] of [Track].
4.
Enter “5” to [Frame Interval].
5.
Enter “2” to [Zoom].
6.
Select “Middle” in [Scan Speed]. (Return to the setting before bleaching.)
7.
Enter the laser intensity value for image acquisition to [Laser Intensity].
8.
Enter the PMT value of each channel to [PMT Voltage].
9.
Enter the image file name to be stored to [Save FileName].
10. Enter a comment in [Comment] if necessary.