7 Acquiring Data
The experiment data before and after photobleaching of certain region can be obtained after
performing an ordinary image acquisition in the first scan, an image acquisition in the Bleach
mode in the second scan , and an ordinary image acquisition in the third scan. When the
REX mode is selected in the third scan, the data of the region other than the specified region
can not be obtained since the laser is irradiated only to the specified region. However, the
REX mode is effective to prevent fading of the obtained data.
1.
Select the [Disabled] option button in the [AOTF] group box in the [Lasers] sub-panel.
2. Click the <XYT> button to start image acquisition.
3. Select the [Bleach mode] option button in the [AOTF] group box in the [Lasers] sub-
panel.
Select the [Bleach mode] option button within 10 seconds before the next
scanning is started.
In the TIME COURSE software (optional), when the REX mask file in the
REX or Bleach mode is changed while image scanning is performed
obtaining the real time graph, the laser is irradiated while the pop-up menu
is displayed. however, the real time graph is not plotted. Please be aware
this before using.
<XYT> button
[Disabled] option button
[Bleach mode] option button
NOTE
NOTE
Содержание Fluoview FV1000
Страница 2: ......
Страница 12: ......
Страница 22: ......
Страница 356: ......
Страница 397: ...APPLIED OPERATIONS Viewing 3D Image OPERATION INSTRUCTIONS 2 3 1 3 Page Fig 2 130 Panel Showing Stereo 3D Images...
Страница 446: ......
Страница 452: ......
Страница 464: ......
Страница 476: ......
Страница 482: ......
Страница 484: ......
Страница 486: ......
Страница 524: ......
Страница 534: ......
Страница 536: ......
Страница 539: ......