APPLIED OPERATIONS
/ Protocol processor
OPERATION INSTRUCTIONS
2 - 1 8 4
Page
One Point!
The protocol to perform the following operations can be assembled after setting the
observation conditions.
z
Specifying the region for bleaching in the laser irradiation mode using REX mask
files.
In Track 4, select a REX mask file in [Mask] of [REX]. Then in the laser irradiation mode
specified by the REX mask file selected, images can be acquired. (When using the system
with AOTF.) To select a REX mask file, display the REX mask file in the image window in
advance.
<The setting example>
For details, see section 2-2-11, “Image Acquisition in the Laser Excitation Mode”.
z
Setting bleaching region by use of rectangular image acquisition (RECT).
Wait 5
sec.
Set Zoom
power to x2
Set Zoom
power to x10
Get 5
images
(
bleach
)
Set Zoom
power to x2
Get one
image
Acquire
image
Save
image
Protocol
Track 1
Track 2
Track 3
Repeat 30 times
Set bleaching
portion
Wait 5
sec.
Acquire
image
Reference image acquisition:
Acquire an image with the Zoom power 2 x in Track 1 in XYT observation.
Photo-bleaching:
Expose laser to photo-bleach with 5 images of acquisition using ZoomIn mode in Track 2 in
XYT observation.
Recovery observation:
Acquire 3 images in XYT observation every 5 seconds with the zoom power 2 x, the field of
view is the same with the reference image, and the images are stored in Track 3.
<Example>
Содержание Fluoview FV1000
Страница 2: ......
Страница 12: ......
Страница 22: ......
Страница 356: ......
Страница 397: ...APPLIED OPERATIONS Viewing 3D Image OPERATION INSTRUCTIONS 2 3 1 3 Page Fig 2 130 Panel Showing Stereo 3D Images...
Страница 446: ......
Страница 452: ......
Страница 464: ......
Страница 476: ......
Страница 482: ......
Страница 484: ......
Страница 486: ......
Страница 524: ......
Страница 534: ......
Страница 536: ......
Страница 539: ......