![NI GPIB-410 Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/ni/gpib-410/gpib-410_user-manual_1685507106.webp)
© National Instruments Corp.
D-1
GPIB-410 User Manual
Appendix D
Common Problems
This appendix contains possible causes for common problems
encountered when operating the GPIB-410.
Problem:
The capture operation does not terminate, even though
enough capture events have occurred to fill up the capture
buffer.
Possible Cause:
•
The
Action
when
buffer
full
field on the Capture Settings
screen is set to
reset
and
continue
. Change this value to
stop
capture
.
•
The value contained in the Interrupt Jumper Setting field on the
Configuration Settings screen does not match the actual jumper
setting on the interface board. Check the hardware configuration
and enter the proper value.
Problem:
There are no highlighted cursors present on the Capture
Display screen.
Possible Cause:
•
The capture buffer is empty. There must be data displayed on the
screen and capturing must be OFF for the cursors to be visible.
•
The contrast on your display is set too low. Adjust the contrast until
the highlighted cursors appear.