GPIB-410
User Manual
March 1994 Edition
Part Number 320053-01
Страница 1: ...Copyright 1986 1994 National Instruments Corporation All Rights Reserved GPIB 410 User Manual March 1994 Edition Part Number 320053 01...
Страница 2: ...ria 0662 435986 Belgium 02 757 00 20 Canada Ontario 519 622 9310 Canada Qu bec 514 694 8521 Denmark 45 76 26 00 Finland 90 527 2321 France 1 48 14 24 24 Germany 089 741 31 30 Italy 02 48301892 Japan 0...
Страница 3: ...shipping costs of returning to the owner parts which are covered by warranty National Instruments believes that the information in this manual is accurate The document has been carefully reviewed for...
Страница 4: ...ted are trademarks or trade names of their respective companies WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS National Instruments products are not designed with componen...
Страница 5: ...2 Hardware and Software Installation 2 1 Step 1 Configure the Hardware 2 1 Step 2 Install the Hardware 2 2 Step 3 Review the Files on the Distribution Diskette 2 4 Step 4 Run the Hardware Diagnostics...
Страница 6: ...en 4 4 Capture Stimulus 4 5 Participate in Handshake 4 7 Buffer Setup Maximum Capacity 4 7 Buffer Setup Action When Buffer Full 4 7 Buffer Setup Auto Write Each Pass 4 8 File Setup Buffer Range for Al...
Страница 7: ...ple 1 Capturing Data From the GPIB A 1 Example 2 Triggering and Pattern Generating A 7 Appendix B Multiline Interface Messages B 1 Appendix C Hardware and Software Configuration C 1 Hardware Revision...
Страница 8: ...410 Interface Board Installed in an IBM Compatible Computer 2 4 Figure 3 1 Monitor Analyzer Main Menu Screen 3 1 Figure 3 2 Monitor Window 3 2 Figure 3 3 Analyzer Window 3 2 Figure 3 4 GPIB Status In...
Страница 9: ...s 2 and 3 Rev B 2 or Higher C 7 Figure C 5 Factory Default IRQ Line 7 Selected Rev B 2 or Higher C 8 Figure C 6 IRQ Lines 6 and 2 Selected Rev B 2 or Higher C 8 Figure C 7 GPIB 410 Parts Locator Diagr...
Страница 10: ...apter 1 Introduction contains a list of the kit contents and optional equipment and the steps for unpacking your GPIB 410 interface board It also contains general information about the GPIB 410 and li...
Страница 11: ...pendix E Using the Pattern Generator describes the Interface Bus Pattern Generating Language IBPGL and syntax conventions for data command strings Appendix F GPIB410 PAT Source Listing contains a sour...
Страница 12: ...drives paths directories programs subprograms subroutines device names functions variables filenames and extensions and for statements and comments taken from program code bold monospace Bold lowerca...
Страница 13: ...ms are used in this manual ANSI American National Standards Institute ASCII American Standard Code for Information Interchange BNC referring to a coaxial connector DMA direct memory access FIFO first...
Страница 14: ...Digital Interface for Programmable Instrumentation Customer Communication We appreciate communicating with the people who use our products We are also very interested in hearing about the application...
Страница 15: ...onal computer into a sophisticated IEEE 488 bus monitor and analyzer The GPIB 410 has low level testing and debugging capabilities as well as high level analysis tools Figure 1 1 GPIB 410 Interface Bo...
Страница 16: ...nterpreted in a high level format You can browse or pattern search through the captured data directly on the screen log it to a disk and obtain a printed output copy With multiple level triggering you...
Страница 17: ...d match the kit parts list given earlier in this chapter Do not remove the board from its plastic bag at this point 2 Your GPIB 410 board is shipped packaged in an antistatic plastic bag to prevent el...
Страница 18: ...at 5 V Temperature Range Operating 0 to 50 C Storage 62 to 71 C Relative Humidity Operating 10 to 90 non condensing Storage 0 to 100 non condensing Maximum Bus Sampling Rate 1 8 MHz Software Specifica...
Страница 19: ...B 410 occupies 32 contiguous bytes of I O space The base address of this 32 byte block of addresses can be any multiple of 20 hex ranging from 000 hex to 3E0 hex The GPIB 410 is configured at the fact...
Страница 20: ...Consult the user or technical reference manual of your personal computer for specific instructions and warnings 1 Turn off your computer and unplug the power cord 2 Remove the top cover or access port...
Страница 21: ...lock input you must locate another open expansion slot for the panel This panel does not require a full length slot 9 Locate and remove the bracket that covers the cut out in the back panel for the sl...
Страница 22: ...s on the Distribution Diskette The GPIB 410 package contains the Distribution Disk for GPIB 410 GPIB Monitor Analyzer Software hereafter referred to as the distribution diskette Verify that your distr...
Страница 23: ...p diskette or in the hard disk directory to run the GPIB 410 software Step 4 Run the Hardware Diagnostics Run the hardware diagnostics by entering the following command diag410 Enter Carefully follow...
Страница 24: ...reens and operation modes and the input conventions used throughout the software To become familiar with the windows and menus of the GPIB 410 software start the GPIB 410 program by entering the follo...
Страница 25: ...of the Monitor Analyzer screen and is visible at all times The analyzer window shown in Figure 3 3 occupies the bottom three quarters of the screen The contents of the analyzer window vary depending...
Страница 26: ...adecimal formats Figure 3 4 GPIB Status Indicators The following list shows how the LEDs and switches appear when they are on and off Switch on Switch off LED on LED off As shown in Figure 3 4 the top...
Страница 27: ...bar The general information display The help error bar The function key line These four areas are illustrated in Figure 3 5 Figure 3 5 Format of the Analyzer Window The title status bar contains the n...
Страница 28: ...reens are selected by pressing F3 F4 F5 F6 or F7 F3 Capture Settings Screen F4 Trigger Specification Screen F5 Capture Display Screen F6 Pattern Generator Screen F7 Configuration Settings Screen Note...
Страница 29: ...r to review and analyze the contents of the capture buffer after the capture has terminated Pattern Generator Screen F6 selects the Pattern Generator screen Use this screen to specify a sequence of pa...
Страница 30: ...at is the arrow keys to highlight the field then press F1 Once the Help screen appears use PgDn or the down arrow key to read the information If you would like to read something that has already scrol...
Страница 31: ...con Alt C Capture Use this mode to capture the status of the GPIB as directed by the parameters defined on the Capture Settings screen Pressing Alt C when a capture is in progress terminates the captu...
Страница 32: ...uence defined on the Trigger Specification screen to occur on the GPIB When this mode is active TRG is displayed in the Title Status bar and T is displayed in the DOS shell icon This mode cannot be en...
Страница 33: ...shell icon This mode cannot be enabled when the Pattern Generator is enabled Alt D DOS Shell Use this function to escape to DOS temporarily without terminating the GPIB410 program All enabled operatio...
Страница 34: ...particular field it must be selected The software is always operating in one of two input modes monitor or analyzer When Monitor Input mode is enabled only those fields in the monitor window can be s...
Страница 35: ...ectness when the field is de selected A selected direct input field is distinguished by the presence of a block cursor superimposed over one of the characters in the field There are two types of direc...
Страница 36: ...you can modify the value of either field by using the indicated keys Changing the value of the parent field causes the sub field of the deactivated option to be erased Key Select Fields These fields p...
Страница 37: ...Overview Chapter 3 GPIB 410 User Manual 3 14 National Instruments Corp SpaceBar Use this key to reset all component fields to their original off or null position...
Страница 38: ...nd the various screens of the Analyzer window and their respective input fields The Monitor Window This window shows the current state of the sixteen GPIB signal lines and indicates which of those lin...
Страница 39: ...ut mode is active indicated by MON in the Title Status bar Press Alt M to enable or disable Monitor Input mode Lines that you assert by setting the corresponding switches to ON remain asserted in betw...
Страница 40: ...red is asserted immediately onto the bus and the Monitor window indicators are adjusted accordingly Only the data lines are affected by this operation no source handshaking is performed Data and Contr...
Страница 41: ...ture operation see Figure 4 2 Figure 4 2 Capture Settings Screen When accessing this screen no input fields can be selected when the following operational modes are enabled Capture Trigger with backgr...
Страница 42: ...uffer To save the contents of the capture buffer to a file on disk use the Capture Write function Alt W To reload the contents of a previously saved file back into the capture buffer use the Capture L...
Страница 43: ...d of an interface clear IFC asserted REN transition Defined as the initial assertion or unassertion of REN SRQ transition Defined as the initial assertion or unassertion of SRQ ATN transition Defined...
Страница 44: ...ake feature is independent of the ACP handshaker Alt A Buffer Setup Maximum Capacity Use this field to allocate a fixed amount of memory in which to store captured bus samples One bus sample requires...
Страница 45: ...yes causes the results of future capture operations to be written to the specified file whenever the capture buffer is full or capturing is stopped To implement a continuous capture operation to disk...
Страница 46: ...e file is written from the capture buffer the existing contents of the file are discarded If the file is loaded into the capture buffer the file must contain valid GPIB 410 capture data Warning The Ca...
Страница 47: ...fied in the pattern sequence Use function keys F8 F9 and F10 to insert append and delete patterns respectively Deleted patterns are pushed back on a stack of unused patterns used by the insert and app...
Страница 48: ...possible the second event will be missed If the selected trigger action is TRIGGER with position TTTTT background capturing may be started when the trigger is armed If so following satisfaction of th...
Страница 49: ...data portion of the pattern and the next row of letters manipulates the control portion Use the Insert key Ins to place a single 8 bit ASCII or 2 digit hexadecimal value directly into the data portion...
Страница 50: ...ion T If the T is placed in the left most position capturing does not begin until the Trigger Spec is satisfied Placing the T in any other position causes background capturing to begin as soon as the...
Страница 51: ...sequence is satisfied The Capture Display Screen This screen shows the current contents of the capture buffer defined on the Capture Settings screen see Figure 4 4 Figure 4 4 Capture Display Screen I...
Страница 52: ...11111 01010 011 UNT The fields columns and indicators of the display window are described in the following paragraphs C This column displays two numeric flags 1 and 2 that label and distinguish the po...
Страница 53: ...on of each buffer entry The left column contains the binary values of the five bus management lines and the right column contains the binary values of the three bus handshake lines The ordering of the...
Страница 54: ...T untalk Appendix B Multiline Interface Messages contains a listing of all IEEE 488 mnemonics and their ASCII equivalents For a complete listing and detailed explanation of all IEEE 488 mnemonics refe...
Страница 55: ...keys are only active when a loaded capture file exceeds the size of the current capture buffer The default displacement for any move or paging operation is one To override this displacement type in a...
Страница 56: ...in the capture buffer The ordering of the signal lines in the pattern is identical to that shown in the Monitor window Construct the search pattern by pressing the indicated letter keys to select a 0...
Страница 57: ...e searching in either direction the GPIB 410 loops around and continues the search from the other end of the buffer If no pattern is found an error message is displayed in the help error bar and the c...
Страница 58: ...during this sequence the switches will contain the last pattern output Direct manipulation of the Monitor switches is not possible while generating is enabled Note During high speed DMA operation you...
Страница 59: ...or window to OFF at the end of a pattern generation sequence There are two options yes clear all switches no leave switches as last set Selecting yes causes the GPIB 410 to stop asserting all lines on...
Страница 60: ...s the GPIB 410 to use commands in a previously prepared file on disk to construct the patterns in the pattern buffer The commands in the file are written in the National Instruments Interface Bus Patt...
Страница 61: ...he following GPIB mnemonics DCL 0x14 PPC 0x05 SPE 0x18 GET 0x08 PPU 0x15 TCT 0x09 GTL 0x01 SDC 0x04 UNL 0x3F LLO 0x11 SPD 0x19 UNT 0x5F NUL 0x00 Alt P Insert an ASCII into the Input Window Alt N Inser...
Страница 62: ...re 4 7 Configuration Settings Screen When accessing this screen no input fields can be selected when the following operational modes are enabled Capture Monitor Input Trigger ACP Handshaker Generate S...
Страница 63: ...or your computer display hardware There are six options Monochrome ROM BIOS Monochrome Synchronized Direct Monochrome High Speed Direct Color ROM BIOS Color Synchronized Direct Color High Speed Direct...
Страница 64: ...verall program performance For most display hardware in use today the High Speed Direct options work satisfactorily Some older video adapters such as the IBM Color Graphics Adapter CGA and its compati...
Страница 65: ...alid port addresses range from 000 to 3E0 hex in multiples of 20 hex default 300 hex A good way to test if the value contained in this field is correct is to examine the sixteen GPIB indicators in the...
Страница 66: ...A channel available on your computer you can select programmed I O PIO operation by setting this field to NONE However because PIO greatly reduces the performance of most capture operations it is reco...
Страница 67: ...already in use by another device The value selected must match the interrupt jumper setting on the interface board itself Warning Attempting to operate some functions of the GPIB 410 software with an...
Страница 68: ...Hz 700 KHz 200 Hz 2 KHz 20 KHz 200 KHz 900 KHz 300 Hz 3 KHz 30 KHz 300 KHz MAXIMUM Selecting one of the numeric options programs the GPIB 410 hardware to regulate the acceptor handshake to an approxim...
Страница 69: ...r configuration variables such as host processor speed and DMA vs PIO Pattern Generator operation the actual source handshake rate observed may be less than requested Help File Name Use this field to...
Страница 70: ...e following screens to be saved to the indicated file upon program exit Capture Settings Screen Trigger Specification Screen Pattern Generator Screen Configuration Settings Screen Press Esc while the...
Страница 71: ...n the Configuration Settings screen The selected keyboard layout type is U S English When the hardware and software are installed step through the following examples Example 1 Capturing Data From the...
Страница 72: ...IB 410 User Manual A 2 National Instruments Corp The Main Menu screen appears as shown in Figure A 1 Figure A 1 Main Screen Notice that the GPIB 410 monitor LEDs are not lit indicating that there is n...
Страница 73: ...icates the maximum number of capture samples that can be stored in PC memory at one time Enter the value 16 For the other fields on the Capture Settings screen the default values are used for this exa...
Страница 74: ...the capturing by pressing Alt C CAP appears on the right side of the title status bar NDAC also becomes asserted on the GPIB 410 monitor because the GPIB 410 is participating in the GPIB handshake as...
Страница 75: ...8 command set refer to Appendix B for a complete listing of the multiline interface messages and their ASCII equivalents At this point the monitor window appears as follows Press k twice for a data tr...
Страница 76: ...data pattern as the Untalk UNT command Refer to Chapter 4 for a description of the fields and functions of the Capture Display screen Use the up arrow key and down arrow key to move the cursors throu...
Страница 77: ...they are acquired 7 Press F2 to return to the Main Menu Example 2 Triggering and Pattern Generating In this example triggering is used with the pattern generator to capture data from the GPIB based on...
Страница 78: ...s size is determined automatically by the GPIB 410 software After the buffer has been loaded the Pattern Generator can be executed any number of times without reloading until a new filename is specifi...
Страница 79: ...ed For this example set the Trigger Spec to trigger when listen address 23 hex is detected as a valid command byte on the GPIB As you may recall from examining GPIB410 PAT this is the address of the i...
Страница 80: ...ccurring one time with no output pulse to be sent to the TRG OUT connector The trigger action specified is TRIGGER with position TTTTT and the trigger point is positioned at the center of the capture...
Страница 81: ...e trigger capture samples into memory If the capture buffer fills up before the trigger occurs the GPIB 410 automatically loops around and reuses the buffer When the trigger pattern is detected on the...
Страница 82: ...The first beep signals the detection of the trigger pattern on the GPIB and you may also notice the TRG indicator change to read CAP Immediately following you hear a second beep which signals the com...
Страница 83: ...ture Display screen refer to Chapter 4 At this point you may wish to experiment by increasing the size of the capture buffer so that more samples are retained either side of the trigger point and then...
Страница 84: ...face message reference list which describes the mnemonics and messages that correspond to the interface functions These multiline interface messages are sent and received with ATN TRUE For more inform...
Страница 85: ...016 14 SO 2E 056 46 MLA14 0F 017 15 SI 2F 057 47 MLA15 10 020 16 DLE 30 060 48 0 MLA16 11 021 17 DC1 LLO 31 061 49 1 MLA17 12 022 18 DC2 32 062 50 2 MLA18 13 023 19 DC3 33 063 51 3 MLA19 14 024 20 DC4...
Страница 86: ...109 m MSA13 PPE 4E 116 78 N MTA14 6E 156 110 n MSA14 PPE 4F 117 79 O MTA15 6F 157 111 o MSA15 PPE 50 120 80 P MTA16 70 160 112 p MSA16 PPD 51 121 81 Q MTA17 71 161 113 q MSA17 PPD 52 122 82 R MTA18 7...
Страница 87: ...of hardware Revision B 2 or higher and Revision C Refer to the appropriate section of this appendix for the revision of hardware you are using Hardware Revision B 2 or Higher This section describes h...
Страница 88: ...Configuration The GPIB 410 is addressed as an I O port and decodes the lower ten address lines of the PC I O channel The base I O address for the GPIB 410 is determined by the switches at position U4...
Страница 89: ...300 to 31F hex on the interface board Note Check to see that this address is not already used by equipment installed in the IBM Personal Computer This address does not conflict with any of the address...
Страница 90: ...arked 1 to select a binary value of 1 for the corresponding address bit Figure C 2 shows the default configuration 300 hex 1 2 3 4 5 6 7 8 9 0 5 4 3 2 1 Off On 1 0 Base Base I O Address Binary Hexadec...
Страница 91: ...XT DMA channel 3 is used by the fixed disk Controller in the IBM Personal Computer XT The software can be configured for programmed I O PIO if no DMA channel is available Configuration jumpers should...
Страница 92: ...the GPIB 410 software Revision B or higher for the pattern generator Refer to the section Hardware Revision C later in this chapter for a discussion of DMA channel configuration for both capturing and...
Страница 93: ...ruments Corp C 7 GPIB 410 User Manual DACK 2 Selected DREQ 2 Selected S1 2 3 1 2 C3 DREQ DACK 1 2 C3 S1 2 3 DACK 3 Selected DREQ 3 Selected S1 2 3 1 2 C3 DREQ DACK 1 2 C3 S1 2 3 Figure C 4 DMA Channel...
Страница 94: ...p posts are used to select the interrupt line The interrupt jumper array W3 is located on the interface board with lines I2 through I7 Figure C 5 shows the factory configuration for IRQ Line 7 Figure...
Страница 95: ...p C 9 GPIB 410 User Manual In the space provided here record the new settings of the base I O address the DMA channel and the IRQ line so that you will have them handy when you install the software GP...
Страница 96: ...e C 7 shows the parts locator diagram with the base I O address switches U65 the DMA channel jumpers and the interrupt jumpers If you change any hardware settings make note of them in the spaces provi...
Страница 97: ...O address value by the GPIB 410 software The GPIB 410 occupies 32 contiguous bytes of I O space The base address of this 32 byte block of addresses can be any multiple of 20 hex ranging from 000 hex...
Страница 98: ...side marked 1 to select a binary value of 1 for the corresponding address bit Figure C 8 shows the default configuration 300 hex 1 2 3 4 5 6 7 8 9 0 5 4 3 2 1 Off On 1 0 Base Base I O Address Binary...
Страница 99: ...he software may be configured for programmed I O PIO if no DMA channel is available Configuration jumpers should remain as they are configured by the factory if PIO is selected Each DMA channel consis...
Страница 100: ...th capturing and sourcing that is input and output Notice that with this configuration DMA may not be used by the capture circuitry and pattern generator at the same time This is the factory default c...
Страница 101: ...S1 2 3 S1 2 3 1 2 C3 DREQ DACK Key Used to set the interrupt selection Figure C 10 DMA Channel 1 for Capturing and DMA Channel 3 for Sourcing Rev C 1 2 C3 S1 2 3 S1 2 3 1 2 C3 DREQ DACK Key Used to s...
Страница 102: ...e 7 One jumper and six dual wire wrap posts are used to select the interrupt line The interrupt jumper array is located on the interface board with lines I2 through I7 Figure C 12 shows the factory co...
Страница 103: ...nnels 2 3 4 5 6 7 INT 2 3 4 5 6 7 INT Figure C 13 IRQ Lines 6 and 2 Selected Rev C In the space provided here record the new settings of the base I O address the DMA channel and the IRQ line so that y...
Страница 104: ...w hardware settings If you have not already done so start up the GPIB410 program by entering the following command gpib410 Enter Press F7 to access the Configuration Settings screen see Figure C 14 Fi...
Страница 105: ...Corp C 19 GPIB 410 User Manual When you exit the GPIB410 program be sure to save the new settings in the default configuration file GPIB410 CFG For more information on the Configuration Settings scre...
Страница 106: ...reset and continue Change this value to stop capture The value contained in the Interrupt Jumper Setting field on the Configuration Settings screen does not match the actual jumper setting on the int...
Страница 107: ...display is set too low Adjust the contrast until the highlighted cursor appears Problem There is no indication that the final trigger pattern was matched there was no beep and the TRG indicator in the...
Страница 108: ...ontroller card connected to a monochrome display Go to the Configuration Settings screen F7 and change the computer display mode field from Color to one of the Monochrome options by using or Press Ent...
Страница 109: ...ppendix D GPIB 410 User Manual D 4 National Instruments Corp There is insufficient memory available to load COMMAND COM Try reducing the size of the capture buffer defined on the Capture Settings scre...
Страница 110: ...e GPIB 410 software is running Using the interactive input window on the Pattern Generator screen you can enter a string of data or command bytes that can then be output to the bus using the source ha...
Страница 111: ...y precede any command The value of n must be an integer constant in the range 1 to 65535 If n is omitted the default repetition factor is 1 The expression command is one of the five command keywords d...
Страница 112: ...tn nrfd ndac assert ATN clear NRFD NDAC b set data assert all data lines c set all atn ren clear all lines assert ATN REN As the last example implies ordering is important when any of the composite re...
Страница 113: ...conceptually similar to set data dav d7 d6 d1 implicit wait for NRFD to be cleared set dav implicit wait for NDAC to be cleared set dav As indicated in the illustration the DAV line is automatically...
Страница 114: ...are rounded to the nearest integral value Timer ticks occur approximately once every 55 msec Examples of PAUSE a pause 50 pause for 50 msec b pause pause for 1 sec c pause 3000 pause for 3 sec 5 comma...
Страница 115: ...g lengthy text strings that cannot fit on one line can be broken down into several shorter HANDSHAKE commands yielding the same results For example the sequence of commands handshake abcd handshake 12...
Страница 116: ...quence Outputting bytes through the input window is roughly equivalent to executing a command file that contains only HANDSHAKE commands Unlike a command file however the input window works in conjunc...
Страница 117: ...UNT 0x5F NUL 0x00 In addition to the above escape sequences two others are provided for entering the ASCII characters and They are as follows Alt P Insert an ASCII into the Input Window Alt N Insert...
Страница 118: ...National Instruments Corp F 1 GPIB 410 User Manual Appendix F GPIB410 PAT Source Listing This information not available in PDF version of this document...
Страница 119: ...GPIB410 PAT Source Listing Appendix F GPIB 410 User Manual F 2 National Instruments Corp This information not available in PDF version of this document...
Страница 120: ...lecting yes in the Auto write each pass field on the screen The Immediate Write function which is activated by pressing Alt W saves the current contents of the capture buffer to the specified file on...
Страница 121: ...PIB410 program use the FMT410 utility The syntax used to start FMT410 is as follows FMT410 flags filename The flags shown in the above line are optional they are described later in this appendix These...
Страница 122: ...e variety of printers only standard ASCII characters are used in FMT410 output For each capture table information is displayed from the associated header followed by the table of saved status entries...
Страница 123: ...0 software only store one capture table per file inumber Begin output range at index number Default range begins at index 0 of each capture table cnumber Limit output range to count of number entries...
Страница 124: ...r conflicting the flag specified last takes precedence There must not be any spaces between a flag and its option s The following examples are possible flag combinations These examples assume that the...
Страница 125: ...Saving and Viewing Capture Information on Disk Appendix G GPIB 410 User Manual G 6 National Instruments Corp If you enter the following command fmt410 xdc Enter the output appears as follows s DEV3 me...
Страница 126: ...l screens and functions operate in the same manner as before with the following important exceptions The Immediate Write function Alt W which was formerly included in the list of functions on the Main...
Страница 127: ...n the Revision C release are as follows Background Operation you can exit to DOS temporarily without terminating the GPIB410 program During DOS operation an optional GPIB 410 icon can be displayed on...
Страница 128: ...entral time In other countries contact the nearest branch office You may fax questions to us at any time Corporate Headquarters 512 795 8248 Technical support fax 800 328 2203 512 794 5678 Branch Offi...
Страница 129: ...ns engineers answer your questions more efficiently If you are using any National Instruments hardware or software products related to this problem include the configuration forms from their user manu...
Страница 130: ...National Instruments software product Version Configuration The problem is List any error messages The following steps will reproduce the problem...
Страница 131: ...s form accurately before contacting National Instruments for technical support helps our applications engineers answer your questions National Instruments Products GPIB 410 Software Revision Number on...
Страница 132: ...on the documentation supplied with our products This information helps us provide quality products to meet your needs Title GPIB 410 User Manual Edition Date March 1994 Part Number 320053 01 Please co...
Страница 133: ...e the errors Thank you for your help Name Title Company Address Phone Mail to Technical Publications National Instruments Corporation 6504 Bridge Point Parkway MS 53 02 Austin TX 78730 5039 Fax to Tec...
Страница 134: ...W Capture Write function File Setup Buffer Range for Alt W field 4 8 to 4 9 File Setup Capture File Name field 4 9 saving capture information on disk 4 5 G 1 Analyzer Window areas of the screen 3 4 Ca...
Страница 135: ...ields columns and indicators 4 15 to 4 17 using the display window 4 18 illustration of 4 14 overview 3 6 selecting from Analyzer Window 3 5 using the search utility 4 19 to 4 20 Capture Echo mode Alt...
Страница 136: ...er Speaker Mode field 4 27 configuration See also Configuration Settings Screen base I O address Revision B 2 or higher C 2 to C 4 Revision C C 11 to C 12 diagnostics for hardware configuration 2 5 DM...
Страница 137: ...ontrol Specification Trigger Pattern field 4 12 Data Line Input character field 4 3 Data Line Input hexadecimal field 4 3 DATA transfer event 4 6 diagnostics for hardware configuration 2 5 direct inpu...
Страница 138: ...ples See tutorial EXTERNAL CLOCK PULSE option 4 6 F File Setup Buffer Range for Alt W field 4 8 to 4 9 File Setup Capture File Name field 4 9 flags for formatting captured data G 4 FMT410 utility G 2...
Страница 139: ...s help error bar Analyzer Window 3 5 Help File Name field 4 32 Help Utility Screen overview 3 7 selecting from Analyzer Window 3 5 I IBPGL See Interface Bus Pattern Generating Language IBPGL IDY respo...
Страница 140: ...Jumper Setting field 4 30 overview 2 1 Revision B 2 or higher C 8 Revision C C 16 to C 17 J jumpers and switches base I O address Revision B 2 or higher C 4 Revision C C 12 DMA channel Revision B 2 or...
Страница 141: ...ion of 3 2 4 1 LEDs and switches 3 3 4 2 Monitor Analyzer Main Menu Screen 3 1 multi pass buffer option 4 8 multiline interface messages B 1 to B 3 N No of Times to Execute field 4 22 numeric fields 3...
Страница 142: ...erview E 1 triggering and pattern generating example A 7 to A 13 Pattern Generator Screen Clear Switches When Done field 4 22 description of 4 20 to 4 24 disabled in certain situations 4 21 discrepanc...
Страница 143: ...1 run time flags G 4 search pattern definition window 4 19 to 4 20 SELECTED GPIB EVENTS option 4 6 Send EOI with Last Byte field 4 22 SET command IBPGL E 2 to E 3 signal lines for DMA channels Revisio...
Страница 144: ...option 4 14 TRIGGER WITH POSITION option 4 13 Trigger mode Alt T 3 9 Trigger Specification Screen considerations for defining 4 11 Data Control Specification Trigger Pattern field 4 12 description of...
Страница 145: ...Index GPIB 410 User Manual Index 12 National Instruments Corp V viewing capture data See saving and viewing capture data...