6.3
Command Descriptions
EC1000SA
6-19
[SOURce:]CURRent:LIMit:AVErage
This sets the upper limit for the output RMS current. The output RMS current limiter setting
ranges from 1.0 A to 10.5 A (for 100 V range) or from 1.0 A to 5.3 A (for 200 V range). To keep
compatible with EC1000S, current average limiter command of EC1000S can be used as RMS
limiter command of EC1000SA.
Syntax:
[SOURce:]CURRent:LIMit:AVErage
<NR2>
Parameter:
1.0 to 10.5 (1.0 to 5.3)
Command example:
CURR:LIM:AVE 5
Query example:
CURR:LIM:AVE ?
Response format:
<NR2>
Response example:
5.0
[SOURce:]CURRent:LIMit:PEAK:HIGH
This sets the upper limit of the output peak current (positive polarity). The output peak current
limiter (positive polarity) setting ranges from 10.0 A to 42.0 A (for 100 V range) or from 5.0 A to
21.0 A (for 200 V range).
Syntax:
[SOURce:]CURRent:LIMit:PEAK:HIGH
<NR2>
Parameters:
10.0 to 42.0 (5.0 to 21.0)
Command example:
CURR:LIM:PEAK:HIGH 10
Query example:
CURR:LIM:PEAK:HIGH?
Response format:
<NR2>
Response example:
10.0
[SOURce:]CURRent:LIMit:PEAK:LOW
This sets the upper limit for output peak current (negative polarity). The output peak current
limiter (negative polarity) setting ranges from
−
42.0 A to
−
10.0 A (for 100 V range) or from
−
21.0
A to
−
5.0 A (for 200 V range).
Syntax:
[SOURce:]CURRent:LIMit:PEAK:LOW
<NR2>
Parameters:
−
42.0 to
−
10.0(
−
21.0 to
−
5.0)
Command example:
CURR:LIM:PEAK:LOW
−
10
Query example:
CURR:LIM:PEAK:LOW?
Response format:
<NR2>
Response example:
−
10.0
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...