6.3
Command Descriptions
EC1000SA
6-44
SYSTem:WRELease
This clears a warning. When a warning is cleared, the SET screen is displayed and key input is
enabled. In order for the warning to be cleared, it is necessary that all the factors of the warning
have been removed and it is ready to be cleared.
If no warning has occurred, or if a warning has occurred but the device is not ready to cancel the
warning, [
−
200, “Execution error”] error occurs.
Syntax:
SYSTem:WRELease
<NR1>
Parameters:
1
Command example:
SYST:WREL 1
Response format:
None
Response example:
None
k) PROGram subsystem
PROGram[:SELected]:EXECute
This controls sequence operations. Control is performed using the STOP (stop), START (run),
HOLD (hold), BRANCH0 (branch 0), or BRANCH1 (branch 1) parameter.
When in AC-EXT, AC-ADD, AC-SYNC, AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC mode,
sequences cannot be executed. ([3, “Invalid in this mode”] error occurs.)
When output is off, sequence execution is disabled. ([2, “Invalid with output off”] error occurs.)
Syntax:
PROGram[:SELected]:EXECute
<CRD>
Parameters:
STOP / START / HOLD / BRANCH0 / BRANCH1
Command example:
PROG:EXEC START
Response format:
None
Response example:
None
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...