6.3
Command Descriptions
EC1000SA
6-29
[SOURce:]VOLTage:ADJust:OFFSet:DC
This sets the DC offset voltage adjustment value in mV when in AC+DC-INT, AC+DC-EXT,
AC+DC-ADD, AC+DC-SYNC mode. It ranges from
−
250.0 mV to 250.0 mV. (The setting range
is common to 100 V range and 200 V range.)
The DC offset voltage adjustment value cannot be set unless the mode is AC-INT, AC-EXT,
AC-ADD, AC-SYNC mode. ([3, “Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]VOLTage:OFFSet:DC
<NR2>
Parameters:
−
250.0 to 250.0
Command example:
VOLT:ADJ:OFFS:DC 10
Query example:
VOLT:ADJ:OFFS:OFFS:DC?
Response format:
<NR2>
Response example:
10.0
[SOURce:]VOLTage:RANGe
This sets the output voltage range. The output voltage range setting is 100 V or 200 V.
However, when output is on, output voltage range cannot be changed. ([1, “Invalid with output
on”] error occurs.)
Syntax:
[SOURce:]VOLTage:RANGe
<NR1>
Parameters:
100/200
Command example:
VOLT:RANG 100
Query example:
VOLT:RANG?
Response format:
<NR1>
Response example:
100