4.4
Using the Sequence Function
EC1000SA
4-43
Setting step 2
10.
Press the
or
key to move the cursor to “STEP”.
11.
Press the
ENTER
key to display the modification box.
12.
Press the
and
keys to move the cursor to 1’s place, then turn the MODIFY dial to set “2”.
The set value will be reflected in output immediately.
13.
Press the
CANCEL
or
ENTER
key to close the modification box.
14.
Press the
key to move the cursor to “TIME”.
15.
Press the
ENTER
key to display the modification box.
16.
Press the
and
keys to move the cursor to the 0.1’s place, then turn the MODIFY dial to set
“2” (this sets “0.2 seconds”). The set value will be reflected in output immediately.
17.
Press the
CANCEL
or
ENTER
key to close the modification box.
18.
Using the same operations as in steps 2 to 5, set DC VOLT = 25, and using the same operations as
in steps 7 to 9, set the DC voltage operation type as “KEEP”.
Use the same operation steps to set step 3 and subsequent steps.