6.3
Command Descriptions
EC1000SA
6-21
[SOURce:]FUNCtion[:SHAPe][:IMMediate]
This sets the output waveform. The output waveform setting is SIN/SQU/ARB1 to ARB16.
When in AC-EXT/AC+DC-EXT mode, the output waveform cannot be set. ([3, “Invalid in this
mode”] error occurs.)
Syntax:
[SOURce:]FUNCtion[:SHAPe][:IMMediate]
<CRD>
Parameters:
SIN/SQU/ARB1 to ARB16
Command example:
FUNC SIN
Query example:
FUNC?
Response format:
<CRD>
Response example:
SIN
[SOURce:]MODE
This sets the output mode. There are eight output modes, each of them is a combination of one
operation mode (AC or AC+DC) and one signal source mode, (INT, EXT, ADD, or SYNC).
Consequently, the output mode setting is AC-INT, AC-EXT, AC-ADD, AC-SYNC, AC+DC-INT,
AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC.
However, when output is on, the output mode cannot be changed. ([1, “Invalid with output on”]
error occurs.)
Syntax:
[SOURce:]MODE
<CRD>
Parameters:
AC-INT/AC-EXT/AC-ADD/AC-SYNC/ACDC-INT
ACDC-EXT/ACDC-ADD/ACDC-SYNC
Command example:
MODE ACDC-INT
Query example:
MODE?
Response format:
<CRD>
Response example:
ACDC-INT
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...