5.4
MEASURE Screen
EC1000SA
5-24
5.4.3 Output power measurement
The output power values described below are always displayed.
a) Effective power
•
Full scale:
1200 W
•
Display resolution:
1 W
•
Measurement accuracy: at 45 Hz to 65 Hz
±
(2% of rdg +1 W)
at DC
±
(3% of rdg +12 W)
(at 23
±
5°C, load with the power factor 0.5 to lower than 1, 50 V or higher
output voltage, output current is 10% to 100% of the maximum output
current)
b) Apparent power
Calculated as output voltage RMS value × output current RMS value
•
Full scale:
1400 VA
•
Display resolution:
1 VA
•
Measurement accuracy: at 45 Hz to 65 Hz
±
(2% of rdg +1 VA)
at DC
±
(3% of rdg +12 VA)
(at 23
±
5°C, 50 V or higher output voltage, output current is 10% to 100%
of the maximum output current)
c) Reactive power
Calculated as
(
)
apparent power 2-
(
)
effective power 2
•
Full scale:
1400 var
•
Display resolution:
1 var
•
Measurement accuracy:
±
(2% of rdg +1 var)
(at 23
±
5°C, load with the power factor 0 to lower than 0.5, 45 Hz to 65 Hz,
50 V or higher output voltage, output current is 10% to 100% of the
maximum output current)
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...