4.2
Measurement of Inrush Current
EC1000SA
4-18
4.
Press the
TOP MENU key or the CANCEL key to return to the SET menu screen.
⇒
The measured value window shows the peak value set in the measurement display.
SEQUENCE
STOP
START
TOP
MENU
LOCAL
HOLD
Peak value
Peak hold value
4.2.4 Reset the peak current hold value
The EC1000SA includes a function that resets the peak current hold value.
A reset sets the peak current hold value to 0 Apk, and the maximum peak value of current is updated
until the next reset is performed.
The current peak value is reset automatically at power-on.
To reset the peak current hold value, select “MEASURE” in the MISC menu
.
Operation steps
1.
Press the
or
key to move the cursor to the
icon, then press the
ENTER
key.
ENTER
CANCEL
Cursor
⇒
The MISC menu is displayed.