4.1
Use as a DC Power Source
EC1000SA
4-5
4.1.3 Setting the output voltage range
Select 100 V or 200 V as the output voltage range.
As is shown in
Table4-2
, the range that can be set for the output voltage differs according to the
selected output voltage range.
Table4-2. Setting Range Options for Various Output Voltage Ranges
To select an output voltage range, select
in the SET menu.
The output voltage range cannot be changed when output is on.
Operation steps
1.
Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor
2.
Press the
ENTER
key to display the selection box.
ENTER
CANCEL
Selection box
Setting
Setting range
100 V range
200 V range
Unit
DC voltage
-220.0 to +220.0
-440.0 to +440.0
V
AC voltage
SIN/SQU
0.0 to 155.0
0.0 to 310.0
Vrms
ARB1 to ARB16
0.0 to 440.0
0.0 to 880.0
Vp-p
Peak current limiter (positive)
+10.0 to +42.0
+5.0 to +21.0
A
Peak current limiter (negative)
-42.0 to -10.0
-21.0 to
-5.0
A
RMS current limiter
1.0 to
10.5
1.0 to
5.3
A
Positive voltage setting limit
+0.1 to +220.0
+0.1 to +440.0
V
Negative voltage setting limit
-220.0 to
-0.1
-440.0 to
-0.1
V
External input gain
0.0 to 220.0
0.0 to 440.0
times
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...